Hide thumbs Also See for PROFESSIONAL HR 500 Plus:
Table of Contents

Advertisement

2. Loosen the thumbscrews on the lamp cover and pull the cover straight out.
3. Push down the lamp release lever.
4. Remove the old lamp from the lamp socket.
5. Fully seat the new lamp in the socket; polarity is not important.
6. With the socket leads facing left, reinstall the lamp in the lamp holder.
7. Reinstall the lamp cover and tighten the thumbscrews.
8. Calibrate the scanner (see Calibrating the Scanner on page 3-5).
May 2002
leads
socket
lamp cover
P192_0023GC
When you remove the cover, the safety switch is released, removing power to
the lamp socket.
CAUTION: Wait at least five minutes for the lamp to cool.
CAUTION: In the next step, to ensure correct alignment with the light bar,
it is extremely important to push the lamp to the back of the
lamp holder with the leads facing to the left. Improper
alignment affects the operation of the scanner.
Power is reapplied to the lamp socket and the lamp should light.
Get other manuals https://www.bkmanuals.com
Maintaining the Equipment
lamp
Hot Surface
symbol
thumbscrews
4-3