Maintaining the Equipment
Cleaning the Light Bar
If artifacts appear on your scanned images, clean the light bar.
IMPORTANT: Wear white cotton gloves for this procedure to prevent oils and dirt
1. Fold a dry, untreated microfiber cleaning cloth so the edge of the cloth is
2. Gently move the edge of the microfiber cloth or e-wipe down the length of the
3. Calibrate the scanner.
4. If an artifact remains after cleaning the light bar, check the light bar with a

Replacement Procedures

Replacing the Lamp
White cotton gloves are required for this procedure.
One spare lamp is provided with the scanner.
IMPORTANT: Use only the GE ELC Lamp with the HR Film Scanner (see
It is not necessary to power down the scanner before changing the lamp.
P192_0016HC
1. Open the lamp-and-filter-access door.
4-2
from being absorbed into the microfiber cleaning cloth.
about the thickness of the light bar.
NOTE: You can use an e-wipe, manufactured by Photographic Solutions, Inc.,
instead of a microfiber cleaning cloth.
light bar one time.
magnifying glass for damage, such as a scratch or a chip. If there is damage
to the light bar, contact your Kodak representative for service.
Appendix B: Ordering Accessories and Supplies).
Get other manuals https://www.bkmanuals.com
lamp-and-filter-
access door
lift door open from here