Test specification for enclosure port immunity to RF wireless communications equipment.
Test frequency
Modulation
Immunity Test Level
(MHz)
(V/
m)
385
Pulse
modulation
27
18 Hz
450
FM
±
5 kHz deviation
28
1kHz sine
710
Pulse
modulation
9
217
Hz
745
780
810
Pulse
modulation
28
18 Hz
870
930
1720
Pulse
modulation
28
217
Hz
1845
1970
2450
Pulse
modulation
28
217
Hz
5240
Pulse modulation
9
217 Hz
5500
5785
NOTE:
If necessary to achieve the
IMMUNITY
TEST
LEVEL,
the distance between
the
transmitting antenna and
the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m, the
1
m test distance is permitted by IEC
61000-4-3.
a).
The
carrier shall be modulated using a 50% duty cycle square wave signal.
b). AS an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used because while it
does not represent actual modulation, it would be worst case.