Viewing SelfCalibration result
Attention
• The result can be viewed when "QC History" is "On" (see
Procedure
1. Select "RadiCS SelfQC" from the Adjustment menu, and press
2. Select "SelfCalibration" from "RadiCS SelfQC", and press
3. Select "Result" with
4. Select "DICOM", "Custom" or "CAL" with
5. Press
to exit.
The SelfCalibration result (Max. Error Rate, Last Time, and Error Code) appears.
4-2. Checking Grayscale
Grayscale Check can be executed and the latest result can be viewed. In addition, when the schedule is set using
RadiCS, the next grayscale check schedule appears.
●
Executing Grayscale Check
Attention
• Do not execute this function when a panel protector is attached.
• Execute Grayscale Check under the actual temperature of the monitor usage environment.
Procedure
1. Select "RadiCS SelfQC" from the Adjustment menu, and press
2. Select "Grayscale Check" from "RadiCS SelfQC", and press
3. Select "Execute" with
4. Press
.
Grayscale Check is executed.
●
Viewing grayscale check results
Attention
• The results can be viewed when "QC History" is "On".
Procedure
1. Select "RadiCS SelfQC" from the Adjustment menu, and press
2. Select "Grayscale Check" from "RadiCS SelfQC", and press
3. Select "Result" with
4. Select "DICOM", "Custom" or "CAL" with
5. Press
to exit.
Up to the five latest grayscale check results (Passed/Failed/Canceled/ERR) appear.
or .
or .
or .
or .
or .
"4-4. Setting On/Off for Warning and QC History" (page
.
.
.
.
.
.
Chapter 4 Controlling Monitor Quality
26)).
25