Environmental
conditions
Aircraft use
Electromagnetic
compatibility
IEC 60601-1
classification
AutoSet CS-A technical specifications
Mode pressure
ranges
Maximum single
fault pressure
8
Operating temperature: +5°C to +35°C
Note: The air flow for breathing produced by this therapy device can be higher
than the temperature of the room. Under extreme ambient temperature
conditions (40ºC) the device remains safe.
Operating humidity: 10 to 95% non-condensing
Operating altitude: Sea level to 2,591 m; air pressure range 1013 hPa to 738 hPa
Storage and transport temperature: -20°C to +60°C
Storage and transport humidity: 10 to 95% non-condensing
ResMed confirms that the device/s meets the Federal Aviation Administration
(FAA) requirements (RTCA/DO-160, section 21, category M) for all phases of air
travel.
Product complies with all applicable electromagnetic compatibility requirements
(EMC) according to IEC60601-1-2, for residential, commercial and light industry
environments.
It is recommended that mobile communication devices are kept at least 1 m
away from the device.
Information regarding the electromagnetic emissions and immunity of this
ResMed device can be found on www.resmed.com, on the Products page
under Service and Support. Click on the PDF file for your language.
Class II (double insulation), Type BF, Ingress protection IP21
CPAP mode
Set Pressure: 4–20 cm H
ASV and ASVAuto mode
EPAP: 4–15 cm H
O; PS: 0–20 cm H
2
Maximum single fault steady state pressure: 30 cm H
for > 6 sec; 40 cm H
O
2
O
2
O—if pressure exceeded for >1 sec
2
O—if pressure exceeded