ResMed S9 AutoSet CS-A Information Manual page 5

Adaptive servo-ventilator
Table of Contents

Advertisement

Problem/Possible cause
You may have removed the
SD card before settings were
copied to the device.
The following message is NOT displayed on the LCD after you try to update the settings using
Settings updated successfully, press any key
the SD card:
The settings were not updated.
Alarms
If the system has not been properly assembled, the device will trigger an alarm. Check that the air
tubing has been properly attached to the device and mask (and humidifier if used).
Problem/Possible cause
Alarm is activated and the LCD screen display disappears
Power failure.
Power cord is disconnected or
mains power switch is turned
off during therapy.
Displays message:
There is excessive leak.
Solution
Reinsert the SD card and wait for the Home screen or the
"Settings updated successfully, press any key"
appear on the LCD.
Note: This message only appears once. If you re-insert the SD card
after you have updated your settings, the message will not be re-
displayed.
Contact your clinician/service provider immediately.
Solution
Remove your mask until power is restored.
Ensure the power cord is connected and the mains power switch
(if available) is on.
High leak, please check system setup and all connections
Adjust position of mask and headgear.
Connect the air tubing firmly at both ends.
message to