Problem/Possible cause
Displays message:
Flow is high because air tubing
is not connected properly.
Note: The tube disconnection
check may not operate when an
antibacterial filter is used.
Displays message:
Air tubing is blocked.
Displays message:
Oximeter sensor is not attached
properly.
Oximeter sensor might be
faulty.
Displays message:
Non-vented mask is used.
Mask vents might be blocked.
A low EPAP in conjunction with
supplemental oxygen may result
in false triggering of this alarm
on a vented mask.
Displays message:
Oximeter adapter is not
attached properly.
6
Solution
No tube, please check your tube is connected
Connect the air tubing firmly at both ends.
Tube blocked, please check your tube
Check your air tubing and remove any blockages. Disconnect the
power cord and then reconnect it to restart the device.
No SpO2 data, check oxi sensor attachment to module/finger
Ensure that the oximeter sensor is attached properly to the module
and the patient's finger.
If the message appears repeatedly but the oximeter is attached
properly to the module and the patient's finger, the oximeter
sensor might be faulty.
Contact your service provider or exchange the oximeter.
Non-vented mask, use vented mask or unblock mask vents
Only use a vented mask.
Check if you have sufficient venting. Unblock mask vents if
necessary.
Talk to your clinician about your settings.
No oximeter, check/connect oximeter adapter
Ensure that the oximeter adapter is attached properly.