Using An External Source; Outputting The Ai Convert Clock Signal - National Instruments PXI-6031E User Manual

Daq
Table of Contents

Advertisement

Chapter 2
Analog Input
E Series User Manual

Using an External Source

You can use a signal connected to any PFI or RTSI <0..6> pin as the source
of ai/ConvertClock. Figure 2-29 shows the timing requirements of the
ai/ConvertClock source.
Rising-Edge
Polarity
Falling-Edge
Polarity
Figure 2-29. ai/ConvertClock Source Timing Requirements

Outputting the AI Convert Clock Signal

You can configure the PFI 2/AI CONV CLK pin to output the
ai/ConvertClock signal. The output pin reflects the ai/ConvertClock signal
regardless of what signal you specify as its source.
Figure 2-30 shows the timing of behavior of the PFI 2/AI CONV CLK pin
configured as an output.
Figure 2-30. PFI 2/AI CONV CLK Timing Behavior
The PFI 2/AI CONV CLK pin is configured as an input by default.
t
w
t
= 10 ns minimum
w
t
w
t
= 50 to 150 ns
w
2-42
ni.com

Advertisement

Table of Contents
loading

This manual is also suitable for:

E series