Calibration Adjustment - Fluke v3000 FC Calibration Manual

Wireless ac voltage
Hide thumbs Also See for v3000 FC:
Table of Contents

Advertisement

v3000 FC
Calibration Manual

Calibration Adjustment

The Product features closed-case calibration adjustment and uses known
reference sources. The Product measures the applied reference source,
calculates correction factors, and stores the correction factors in nonvolatile
memory.
Should the Product fail any of the performance tests, do the calibration
adjustment procedure.
When "C-01" is shown on the display, apply the correct input signal shown in
Table 4 to the Product. Then push  to confirm the calibration step. If the input
signal does not satisfy the calibration requirement, "Err" is shown. If the signal is
not stable, it can be necessary to push  several times to confirm the calibration.
After confirmation, the Product goes to the subsequent calibration step.
After you push , wait until the calibration step number advances
before you change the Calibrator source. Some adjustment steps
can take several seconds to execute before the Product goes to the
subsequent step.
Set the calibrator to Standby after you complete adjustment of each
function.
Input each signal to the Product in the sequence shown in Table 4. When the last
calibration point is recorded, "End" shows on the display.
While the calibration adjustment points are shown in Table 4, the
Product also can show the necessary inputs. For each step, push
 to see the necessary voltage input. The necessary frequency for
all steps is 55 Hz.
C-01
C-02
C-03
C-04
C-05
10
Table 4. Calibration Adjustment
Calibration Step
Note
Note
Calibrator Output Signal
60 mV ac, 55 Hz
600 mV ac, 55 Hz
6 V ac, 55 Hz
60 V ac, 55 Hz
600 V ac, 55 Hz

Advertisement

Table of Contents
loading

Table of Contents