Backlight

To verify that the backlight functions:
1. With the Product on, push .
2. The backlight will come on. If it does not, repair is necessary. See "Contact

Keypad Test

To verify that the keypad functions, turn on the Product and push each button
separately. Each button push will turn on a display annunciator. If the buttons do
nothing, repair is necessary. See "Contact Fluke".

AC Voltage Test

Before you do the ac voltage test:
1. Make sure that you have the necessary equipment. See Table 2.
2. Make sure the Product battery is good and replace it if necessary. See
3. Warm up the Calibrator as necessary. Refer to its specifications.
4. Let the temperature of the unit under test (UUT) become stable to room
To do the ac voltage test:
1. Connect the Product to Calibrator COM and VAC jacks. See Figure 2.
2. Apply the input level for each step shown in Table 3.
3. Compare the indication on the Product display with the UUT reading limits in
4. If the display indication falls outside of the range shown in Table 3, calibration
Test
AC Volts
6
Fluke".
"Battery Replacement".
temperature.
Table 3.
adjustment or repair of the Product is necessary. See "Calibration
Adjustment".
Table 3. Performance Tests
Calibrator
Resolution
Output
0.060 V, 45 Hz
0.001 V
0.600 V, 500 Hz
0.001 V
6.000 V, 1000 Hz
0.001 V
35.00 V, 45 Hz
0.01 V
60.00 V, 1000 Hz
0.01 V
75.0 V, 500 Hz
0.1 V
600.0 V, 1000 Hz
0.1 V
1000 V, 1000 Hz
1 V
Specification
Low
±[1.0 % + 3]
0.056
±[1.0 % + 3]
0.591
±[2.0 % + 3]
5.877
±[1.0 % + 3]
34.62
±[2.0 % + 3]
58.77
±[1.0 % + 3]
74.0
±[2.0 % + 3]
587.7
±[2.0 % + 3]
977
High
0.064
0.609
6.123
35.38
61.23
76.0
612.3
1023