Sample Period; Duplex Mode; Linefeed - Fluke 7381 User Manual

Calibration bath
Table of Contents

Advertisement

Sample Period

Sample period is the next parameter in the Serial Interface Parameter menu. The sample period
is the time, in seconds, between temperature measurements transmitted from the serial
interface. If you set the sample rate (
over the serial interface approximately every 5 seconds. A sample period of 0 disables the
automatic sampling.
1. Push SET to choose to set the sample period. The current sample period shows.
=
SA
1
2. Push UP or DOWN to adjust the value.
3. Push SET to store the sample rate.

Duplex Mode

Duplex mode (
dUPL
) is the next parameter in the Serial Interface Parameter menu. Set the duplex
mode to full duplex or half duplex. With full duplex, any commands received by the Bath from the
serial interface are immediately echoed or transmitted back to the device of origin and the
command executes. With half duplex, the commands are executed but not echoed.
To change duplex mode:
1. Push SET to access the mode setting.
=
dUP
FULL
2. Push UP or DOWN to change the mode (
3. Push SET.

Linefeed

Linefeed mode is the final parameter in the Serial Interface Parameter. Linefeed mode (
enables (
On
) or disables (
any carriage-return. Push SET to access the linefeed parameter.
To change linefeed mode:
1. Push SET to access the mode setting.
2. Push UP or DOWN to change the mode.
3. Push SET.
SAnPLE
) to 5, the Bath transmits the current measurement
Current sample period (seconds)
Current duplex mode setting
HALF
OFF
) linefeed character (LF, ASCII 10) transmission after transmission of
or
FULL
).
Calibration Bath
Controller Operation
LF
)
39