Controller Configuration; Probe Parameters; Alpha - Fluke 7381 User Manual

Calibration bath
Table of Contents

Advertisement

CO= 110C
1. Push UP or DOWN to change the cutout set-point.
DOWN
CO= 75C
2. Push SET to accept the new cutout set-point.
The next function is the configuration menu. Push EXIT to show the Bath temperature.

Controller Configuration

Configuration and operating options and calibration parameters follow the cutout set-point
function in the secondary menu. Enter the secondary menu and push SET until the display shows
COnFIG
. Push SET again to enter the configuration menu. The configuration menu has five sets of
configuration parameters:

probe parameters

operating parameters
serial interface parameters
IEEE-488 interface parameters
controller calibration parameters
Push UP or DOWN to highlight the desired parameter, and push SET.
Probe Parameters
Push SET to enter the menu. The probe parameters menu (
and ALPHA, which characterize the resistance-temperature relationship of the platinum control
probe. Adjust these parameters if necessary to improve Bath accuracy, see
1. Push SET to access the probe parameters.
2. Push UP or DOWN to change the value of the parameter.
3. After you reach the desired value, push SET to store the parameter as the new value. Push
EXIT to escape the parameter and ignore any changes made.
R
0
This probe parameter, normally set for 100.000 Ω , refers to the resistance of the control probe at
0 °C.

ALPHA

This probe parameter, normally set for 0.00385°C
between 0 °C and 100 °C.
Cutout set-point
Decrement display
New cutout set-point
PrObE
) contains the parameters, R0
–1
, refers to the average probe sensitivity
Calibration Bath
Controller Operation
Cooling
Control.
35