Endress+Hauser Proline Cubemass C 100 Operating Instructions Manual

Endress+Hauser Proline Cubemass C 100 Operating Instructions Manual

Coriolis flowmeter
Hide thumbs Also See for Proline Cubemass C 100:
Table of Contents

Advertisement

Quick Links

BA01183D/06/EN/03.20
71479739
2020-06-01
Valid as of version
01.02.zz (Device firmware)
Products
Operating Instructions
Proline Cubemass C 100
Ethernet/IP
Coriolis flowmeter
Solutions
Services

Advertisement

Table of Contents
loading

Summary of Contents for Endress+Hauser Proline Cubemass C 100

  • Page 1 Products Solutions Services BA01183D/06/EN/03.20 71479739 2020-06-01 Valid as of version 01.02.zz (Device firmware) Operating Instructions Proline Cubemass C 100 Ethernet/IP Coriolis flowmeter...
  • Page 2 • The manufacturer reserves the right to modify technical data without prior notice. Your Endress+Hauser Sales Center will supply you with current information and updates to these instructions. Endress+Hauser...
  • Page 3: Table Of Contents

    Proline Cubemass C 100 Ethernet/IP Table of contents Table of contents Mounting the measuring device ... 24 About this document ....6 6.2.1...
  • Page 4 ......70 13.3 Endress+Hauser services ....96 Operation .
  • Page 5 Proline Cubemass C 100 Ethernet/IP Table of contents 16.2 Function and system design ... . . 101 16.3 Input ......102 16.4 Output .
  • Page 6: About This Document

    About this document Proline Cubemass C 100 Ethernet/IP About this document Document function These Operating Instructions contain all the information that is required in various phases of the life cycle of the device: from product identification, incoming acceptance and storage, to mounting, connection, operation and commissioning through to troubleshooting, maintenance and disposal.
  • Page 7: Symbols For Certain Types Of Information

    • W@M Device Viewer (www.endress.com/deviceviewer): Enter the serial number from nameplate • Endress+Hauser Operations App: Enter the serial number from the nameplate or scan the 2D matrix code (QR code) on the nameplate Detailed list of the individual documents along with the documentation code →...
  • Page 8: Standard Documentation

    About this document Proline Cubemass C 100 Ethernet/IP 1.3.1 Standard documentation Document type Purpose and content of the document Technical Information Planning aid for your device The document contains all the technical data on the device and provides an overview of the accessories and other products that can be ordered for the device.
  • Page 9: Safety Instructions

    Proline Cubemass C 100 Ethernet/IP Safety instructions Safety instructions Requirements for the personnel The personnel for installation, commissioning, diagnostics and maintenance must fulfill the following requirements: ‣ Trained, qualified specialists must have a relevant qualification for this specific function and task.
  • Page 10: Workplace Safety

    Verification for borderline cases: ‣ For special fluids and fluids for cleaning, Endress+Hauser is glad to provide assistance in verifying the corrosion resistance of fluid-wetted materials, but does not accept any warranty or liability as minute changes in the temperature, concentration or level of contamination in the process can alter the corrosion resistance properties.
  • Page 11: Product Safety

    It meets general safety standards and legal requirements. It also complies with the EU directives listed in the device-specific EU Declaration of Conformity. Endress+Hauser confirms this by affixing the CE mark to the device.
  • Page 12: Product Description

    Product description Proline Cubemass C 100 Ethernet/IP Product description The device consists of a transmitter and a sensor. The device is available as a compact version: The transmitter and sensor form a mechanical unit. Product design 3.1.1 Device version with EtherNet/IP communication protocol A0023153 ...
  • Page 13: Incoming Acceptance And Product

    Is the envelope present with accompanying documents? • If one of the conditions is not satisfied, contact your Endress+Hauser Sales Center. • Depending on the device version, the CD-ROM might not be part of the delivery! The Technical Documentation is available via the Internet or via the Endress+Hauser Operations App, see the "Product identification"...
  • Page 14: Transmitter Nameplate

     8 sections • The W@M Device Viewer: enter the serial number from the nameplate (www.endress.com/deviceviewer) • The Endress+Hauser Operations App: Enter the serial number from the nameplate or scan the 2-D matrix code (QR code) on the nameplate. 4.2.1...
  • Page 15: Sensor Nameplate

    Proline Cubemass C 100 Ethernet/IP Incoming acceptance and product identification 4.2.2 Sensor nameplate 3 4 5 6 Order code: Ser. no.: Ext. ord. cd.: Size: Ptest = Material: 0044 Patents Date: 322541-0000 13 14 A0013907  3 Example of a sensor nameplate...
  • Page 16: Symbols On Measuring Device

    Incoming acceptance and product identification Proline Cubemass C 100 Ethernet/IP 4.2.3 Symbols on measuring device Symbol Meaning WARNING! This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in serious or fatal injury. Reference to documentation Refers to the corresponding device documentation.
  • Page 17: Storage And Transport

    Proline Cubemass C 100 Ethernet/IP Storage and transport Storage and transport Storage conditions Observe the following notes for storage: ‣ Store in the original packaging to ensure protection from shock. ‣ Do not remove protective covers or protective caps installed on process connections.
  • Page 18: Measuring Devices With Lifting Lugs

    Storage and transport Proline Cubemass C 100 Ethernet/IP 5.2.2 Measuring devices with lifting lugs CAUTION Special transportation instructions for devices with lifting lugs ‣ Only use the lifting lugs fitted on the device or flanges to transport the device. ‣...
  • Page 19: Installation

    Proline Cubemass C 100 Ethernet/IP Installation Installation Installation conditions 6.1.1 Mounting position Mounting location A0028772 To prevent measuring errors arising from accumulation of gas bubbles in the measuring tube, avoid the following mounting locations in the pipe: • Highest point of a pipeline.
  • Page 20 Installation Proline Cubemass C 100 Ethernet/IP Ø orifice plate, pipe restriction [mm] [in] [mm] [in] ¹⁄₂₄ 0.03 ¹⁄₁₂ 0.06 ¹⁄₈ 0.12 ¹⁄₄ 0.20 Orientation The direction of the arrow on the sensor nameplate helps you to install the sensor according to the flow direction (direction of medium flow through the piping).
  • Page 21: Environmental And Process Requirements

    Proline Cubemass C 100 Ethernet/IP Installation 6.1.2 Environmental and process requirements Ambient temperature range Measuring device • –40 to +60 °C (–40 to +140 °F) • Order code for "Test, certificate", option JM: –50 to +60 °C (–58 to +140 °F) ‣...
  • Page 22: Special Mounting Instructions

    Installation Proline Cubemass C 100 Ethernet/IP Heating NOTICE Electronics can overheat due to elevated ambient temperature! ‣ Observe maximum permitted ambient temperature for the transmitter . ‣ Depending on the fluid temperature, take the device orientation requirements into account .
  • Page 23 Proline Cubemass C 100 Ethernet/IP Installation RUPTURE DISK A0029940 Rupture disk label Wall mounting WARNING Incorrect sensor mounting Risk of injury if measuring tube breaks ‣ The sensor should never be installed in a pipe in a way that it is freely suspended ‣...
  • Page 24: Mounting The Measuring Device

    Installation Proline Cubemass C 100 Ethernet/IP 8 mm A0019768  6 Mounting kit for Cubemass mounting plate 1 x Cubemass mounting plate 4 x screw M5 x 8 4 x grommet Zero point adjustment All measuring devices are calibrated in accordance with state-of-the-art technology.
  • Page 25: Turning The Display Module

    Proline Cubemass C 100 Ethernet/IP Installation 1. Ensure that the direction of the arrow on the nameplate of the sensor matches the flow direction of the fluid. 2. Install the measuring device or turn the transmitter housing so that the cable entries do not point upwards.
  • Page 26: Post-Installation Check

    Installation Proline Cubemass C 100 Ethernet/IP Post-installation check Is the device undamaged (visual inspection)?  Does the measuring device conform to the measuring point specifications? For example: • Process temperature →  112  • Process pressure (refer to the section on "Pressure-temperature ratings" in the "Technical Information"...
  • Page 27: Electrical Connection

    Proline Cubemass C 100 Ethernet/IP Electrical connection Electrical connection NOTICE The measuring device does not have an internal circuit breaker. ‣ For this reason, assign the measuring device a switch or power-circuit breaker so that the power supply line can be easily disconnected from the mains.
  • Page 28: Terminal Assignment

    Electrical connection Proline Cubemass C 100 Ethernet/IP 7.2.3 Terminal assignment Transmitter EtherNet/IP connection version Order code for "Output", option N Depending on the housing version, the transmitters can be ordered with terminals or device plugs. Connection methods available Order code...
  • Page 29: Pin Assignment, Device Plug

    Proline Cubemass C 100 Ethernet/IP Electrical connection 7.2.4 Pin assignment, device plug Supply voltage Assignment DC 24 V Not assigned Not assigned DC 24 V Grounding/shielding A0016809 Coding Plug/socket Plug Device plug for signal transmission (device side) Assignment Coding Plug/socket...
  • Page 30: Connecting The Transmitter

    Electrical connection Proline Cubemass C 100 Ethernet/IP 7.3.1 Connecting the transmitter The connection of the transmitter depends on the following order codes: • Housing version: compact or ultra-compact • Connection version: device plug or terminals A0016924  8 Housing versions and connection versions...
  • Page 31: Ensuring Potential Equalization

    Proline Cubemass C 100 Ethernet/IP Electrical connection For device version with device plug: follow step 6 only. 1. Depending on the housing version, loosen the securing clamp or fixing screw of the housing cover. 2. Depending on the housing version, unscrew or open the housing cover and disconnect the local display from the main electronics module where necessary →...
  • Page 32: Hardware Settings

    Electrical connection Proline Cubemass C 100 Ethernet/IP Hardware settings 7.5.1 Setting the device address EtherNet/IP The IP address of the measuring device can be configured for the network via DIP switches. Addressing data IP address and configuration options 1st octet...
  • Page 33: Ensuring The Degree Of Protection

    Proline Cubemass C 100 Ethernet/IP Electrical connection Ensuring the degree of protection The measuring device fulfills all the requirements for degree of protection IP66/67, Type 4X enclosure. To guarantee degree of protection IP66/67, Type 4X enclosure, carry out the following steps after the electrical connection: 1.
  • Page 34: Operation Options

    Operation options Proline Cubemass C 100 Ethernet/IP Operation options Overview of operating options A0017760 Computer with Web browser (e.g. Internet Explorer) or with "FieldCare" operating tool Automation system, e.g. "RSLogix" (Rockwell Automation) and work station for measuring device operation with Add-on Profile Level 3 for "RSLogix 5000" software (Rockwell Automation)
  • Page 35: Structure And Function Of The Operating Menu

    Proline Cubemass C 100 Ethernet/IP Operation options Structure and function of the operating menu 8.2.1 Structure of the operating menu For an overview of the operating menu for experts: "Description of Device Parameters" document supplied with the device→  121...
  • Page 36: Operating Philosophy

    Operation options Proline Cubemass C 100 Ethernet/IP 8.2.2 Operating philosophy The individual parts of the operating menu are assigned to certain user roles (operator, maintenance etc.). Each user role contains typical tasks within the device lifecycle. Menu/parameter User role and tasks...
  • Page 37: Displaying The Measured Values Via The Local Display (Optionally Available)

    Proline Cubemass C 100 Ethernet/IP Operation options Displaying the measured values via the local display (optionally available) 8.3.1 Operational display The local display is optionally available: Order code for "Display; operation", option B ""4-line, illuminated; via communication". X X X X X X X 1120.50...
  • Page 38: User Roles And Related Access Authorization

    Operation options Proline Cubemass C 100 Ethernet/IP • Density • Reference density Temperature Totalizer  The measurement channel number indicates which of the three totalizers is displayed. Output  Measurement channel numbers Symbol Meaning Measurement channel 1 to 4 The measurement channel number is displayed only if more than one channel is present for the same measured variable type (e.g.
  • Page 39: Access To The Operating Menu Via The Web Browser

    Proline Cubemass C 100 Ethernet/IP Operation options Access authorization to parameters: "Operator" user role Access code status Read access Write access After an access code has been defined.  –– Despite the defined access code, certain parameters can always be modified and thus are excepted from the write protection, as they do not affect the measurement.
  • Page 40: Establishing A Connection

    Operation options Proline Cubemass C 100 Ethernet/IP Network connections Only the active network connections to the measuring device should be used. Switch off all other network connections. In the event of connection problems: Measuring device: Via CDI-RJ45 service interface Device...
  • Page 41 Proline Cubemass C 100 Ethernet/IP Operation options The measuring device works with the Dynamic Host Configuration Protocol (DHCP), on leaving the factory, i.e. the IP address of the measuring device is automatically assigned by the automation system (DHCP server). To establish a network connection via the service interface (CDI-RJ45): set the "Default IP address"...
  • Page 42: Logging On

    Operation options Proline Cubemass C 100 Ethernet/IP 2. Enter the IP address of the Web server in the address line of the Web browser: 192.168.1.212  The login page appears. 2 3 4 A0029417 Picture of device Device name Device tag...
  • Page 43: User Interface

    Proline Cubemass C 100 Ethernet/IP Operation options 8.4.5 User interface A0032879 Picture of device Device name Device tag Status signal Current measured values Navigation area Local display language Header The following information appears in the header: • Device name • Device tag •...
  • Page 44: Disabling The Web Server

    Operation options Proline Cubemass C 100 Ethernet/IP Functions Meaning Configuration and checking of all the parameters required for establishing the connection to Network the measuring device: configuration • Network settings (e.g. IP address, MAC address) • Device information (e.g. serial number, firmware version)
  • Page 45: Access To The Operating Menu Via The Operating Tool

    Proline Cubemass C 100 Ethernet/IP Operation options 1. Select the Logout entry in the function row.  The home page with the Login box appears. 2. Close the Web browser. 3. If no longer needed: Reset modified properties of the Internet protocol (TCP/IP) →  40.
  • Page 46: Fieldcare

    Operation options Proline Cubemass C 100 Ethernet/IP Via service interface (CDI) A0014019 Service interface (CDI = Endress+Hauser Common Data Interface) of the measuring device Commubox FXA291 Computer with FieldCare operating tool with COM DTM CDI Communication FXA291 Via service interface (CDI-RJ45)
  • Page 47 Proline Cubemass C 100 Ethernet/IP Operation options Typical functions: • Configuring parameters of transmitters • Loading and saving device data (upload/download) • Documentation of the measuring point • Visualization of the measured value memory (line recorder) and event logbook For additional information about FieldCare, see Operating Instructions BA00027S...
  • Page 48: Devicecare

    DeviceCare Function scope Tool to connect and configure Endress+Hauser field devices. The fastest way to configure Endress+Hauser field devices is with the dedicated "DeviceCare" tool. Together with the device type managers (DTMs) it presents a convenient, comprehensive solution. For details, see Innovation Brochure IN01047S Source for device description files See information →...
  • Page 49: System Integration

    Proline Cubemass C 100 Ethernet/IP System integration System integration Overview of device description files 9.1.1 Current version data for the device Firmware version 01.02.zz • On the title page of the Operating Instructions • On the transmitter nameplate • Firmware version Diagnostics →...
  • Page 50: Integrating The Measuring Device In The

    System integration Proline Cubemass C 100 Ethernet/IP Integrating the measuring device in the system A detailed description of how to integrate the device into an automation system (e.g. from Rockwell Automation) is available as a separate document: www.endress.com → Select country → Automation → Digital communication → Fieldbus device integration →...
  • Page 51 Proline Cubemass C 100 Ethernet/IP System integration Configuration 4: Input Only Multicast Input Assembly Configurable Instance Size [byte] Min. RPI (ms) Input Assembly Configurable Configuration 0 x 68 – Output Assembly Fix O → T Configuration 0 x C7 –...
  • Page 52 System integration Proline Cubemass C 100 Ethernet/IP Designation Description Byte 10. Totalizer 2 37-40 11. Totalizer 3 41-44 Structure: Code, number, description (e.g.: 16777265 F882 input signal) Detailed description: • Diagnostic information (Verweisziel existiert nicht, aber @y.link.required=' t rue' ) •...
  • Page 53 Proline Cubemass C 100 Ethernet/IP System integration Designation Description (format) Byte Value 8. Not used – 9. Not used – 10. Control totalizer 1 (integer) • -32226: Add • -32490: Reset and stop • -32228: Default value and stop • 198: Reset and add •...
  • Page 54: Commissioning

    Commissioning Proline Cubemass C 100 Ethernet/IP Commissioning 10.1 Function check Before commissioning the measuring device: ‣ Make sure that the post-installation and post-connection checks have been performed. • "Post-installation check" checklist→  26 • "Post-connection check" checklist→  33 10.2 Configuring the device address via software In the "Communication"...
  • Page 55: Defining The Tag Name

    Proline Cubemass C 100 Ethernet/IP Commissioning ‣ Communication →  59 ‣ Low flow cut off →  62 ‣ Partially filled pipe detection →  63 ‣ Advanced setup →  64 10.4.1 Defining the tag name To enable fast identification of the measuring point within the system, you can enter a unique designation using the Device tag parameter and thus change the factory setting.
  • Page 56 Commissioning Proline Cubemass C 100 Ethernet/IP Density unit →  56 Reference density unit →  56 Temperature unit →  57 Pressure unit →  57 Parameter overview with brief description Parameter Description Selection Factory setting Mass flow unit Select mass flow unit.
  • Page 57 Proline Cubemass C 100 Ethernet/IP Commissioning Parameter Description Selection Factory setting Temperature unit Select temperature unit. Unit choose list Country-specific: • °C Result • °F The selected unit applies for: • Electronic temperature parameter (6053) • Maximum value parameter (6051) •...
  • Page 58: Selecting And Setting The Medium

    Commissioning Proline Cubemass C 100 Ethernet/IP 10.4.3 Selecting and setting the medium The Select medium wizard submenu contains parameters that must be configured in order to select and set the medium. Navigation "Setup" menu → Medium selection ‣ Medium selection →...
  • Page 59: Configuring The Communication

    Proline Cubemass C 100 Ethernet/IP Commissioning Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry Select medium – Select medium type. • Liquid Liquid • Gas Select gas type The Gas option is selected in Select measured gas type.
  • Page 60 Commissioning Proline Cubemass C 100 Ethernet/IP Navigation "Setup" menu → Communication ‣ Communication →  60 MAC address →  60 Default network settings DHCP client →  60 IP address →  60 Subnet mask →  61 Default gateway →...
  • Page 61 Proline Cubemass C 100 Ethernet/IP Commissioning Parameter Description User interface / Selection /  Factory setting User entry Subnet mask Displays the subnet mask. 4 octet: 0 to 255 (in the 255.255.255.0 particular octet) If the DHCP client is switched off and write access is enabled, the Subnet mask can also be entered.
  • Page 62: Configuring The Low Flow Cut Off

    Commissioning Proline Cubemass C 100 Ethernet/IP 10.4.5 Configuring the low flow cut off The Low flow cut off submenu contains the parameters that must be set in order to configure the low flow cut off. Navigation "Setup" menu → Low flow cut off ‣...
  • Page 63: Configuring The Partial Filled Pipe

    Proline Cubemass C 100 Ethernet/IP Commissioning 10.4.6 Configuring the partial filled pipe detection The Partially filled pipe detection submenu contains parameters that have to be set for configuring empty pipe detection. Navigation "Setup" menu → Partially filled pipe detection ‣...
  • Page 64: Advanced Settings

    Commissioning Proline Cubemass C 100 Ethernet/IP 10.5 Advanced settings The Advanced setup submenu together with its submenus contains parameters for specific settings. The number of submenus can vary depending on the device version, e.g. viscosity is available only with the Promass I.
  • Page 65 Proline Cubemass C 100 Ethernet/IP Commissioning Navigation "Setup" menu → Advanced setup → Calculated values ‣ Calculated values ‣ Corrected volume flow calculation →  65 Corrected volume flow calculation External reference density →  65 Fixed reference density →  65 Reference temperature →...
  • Page 66: Carrying Out A Sensor Adjustment

    Commissioning Proline Cubemass C 100 Ethernet/IP Parameter Prerequisite Description Selection / User Factory setting interface / User entry Linear expansion coefficient The Calculated reference Enter linear, medium-specific Signed floating-point density option is selected in expansion coefficient for number the Corrected volume flow calculating the reference calculation parameter density.
  • Page 67: Configuring The Totalizer

    Proline Cubemass C 100 Ethernet/IP Commissioning Parameter overview with brief description Parameter Description Selection / User interface Factory setting Zero point adjustment control Start zero point adjustment. • Cancel Cancel • Busy • Zero point adjust failure • Start Progress Shows the progress of the process.
  • Page 68: Simulation

    Commissioning Proline Cubemass C 100 Ethernet/IP Navigation "Setup" menu → Advanced setup → Administration ‣ Administration →  68 Define access code →  68 Device reset Parameter overview with brief description Parameter Description User entry / Selection Factory setting Define access code...
  • Page 69: Protecting Settings From Unauthorized

    Proline Cubemass C 100 Ethernet/IP Commissioning Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry Assign simulation process variable – Select a process variable for • Off the simulation process that is • Mass flow activated. • Volume flow •...
  • Page 70: Write Protection Via Write Protection

    Commissioning Proline Cubemass C 100 Ethernet/IP 2. Define a max. 16-digit numeric code as an access code. 3. Enter the access code again in the to confirm the code.  The Web browser switches to the login page. If no action is performed for 10 minutes, the Web browser automatically returns to the login page.
  • Page 71: Operation

    Proline Cubemass C 100 Ethernet/IP Operation Operation 11.1 Read out and modify current Ethernet settings If the Ethernet settings such as the IP address of the measuring device are unknown, they can be read out and modified as explained in the following example for an IP address.
  • Page 72: Adjusting The Operating Language

    Operation Proline Cubemass C 100 Ethernet/IP Navigation "Operation" menu → Locking status Function scope of "Locking status" parameter Options Description Hardware locked The write protection switch (DIP switch) for hardware locking is activated on the I/O electronic module. This prevents write access to the parameters .
  • Page 73 Proline Cubemass C 100 Ethernet/IP Operation Reference density →  73 Temperature →  73 Pressure value →  73 Concentration →  74 →  74 Target mass flow Carrier mass flow →  74 Parameter overview with brief description...
  • Page 74: Totalizer" Submenu

    Operation Proline Cubemass C 100 Ethernet/IP Parameter Prerequisite Description User interface Concentration For the following order code: Displays the concentration that is Signed floating-point Order code for "Application package", currently calculated. number option ED "Concentration" Dependency  The software options currently...
  • Page 75: Adapting The Measuring Device To The Process

    Proline Cubemass C 100 Ethernet/IP Operation Parameter overview with brief description Parameter Prerequisite Description User interface Totalizer value 1 to n One of the following options is selected Displays the current totalizer counter Signed floating-point in the Assign process variable value.
  • Page 76: Function Scope Of The "Control Totalizer" Parameter

    Operation Proline Cubemass C 100 Ethernet/IP Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry Control Totalizer 1 to n A process variable is selected Control totalizer value. • Totalize Totalize in the Assign process variable • Reset + hold parameter (→...
  • Page 77: Diagnostics And Troubleshooting

    Proline Cubemass C 100 Ethernet/IP Diagnostics and troubleshooting Diagnostics and troubleshooting 12.1 General troubleshooting For local display Problem Possible causes Remedial action Local display dark and no output Supply voltage does not match that Apply the correct supply voltage signals specified on the nameplate.
  • Page 78 Diagnostics and troubleshooting Proline Cubemass C 100 Ethernet/IP Problem Possible causes Remedial action Not connecting to Web server • Incorrect IP address 1. If addressing via hardware: open • IP address is not known the transmitter and check the IP address configured (last octet).
  • Page 79: Diagnostic Information Via Light Emitting

    Proline Cubemass C 100 Ethernet/IP Diagnostics and troubleshooting 12.2 Diagnostic information via light emitting diodes 12.2.1 Transmitter Various light emitting diodes (LEDs) on the main electronics module of the transmitter provide information on device status. Color Meaning Power Supply voltage is off or too low...
  • Page 80: Diagnostic Information On Local Display

    Diagnostics and troubleshooting Proline Cubemass C 100 Ethernet/IP 12.3 Diagnostic information on local display 12.3.1 Diagnostic message Faults detected by the self-monitoring system of the measuring device are displayed as a diagnostic message in alternation with the operational display. Operational display in alarm condition...
  • Page 81 Proline Cubemass C 100 Ethernet/IP Diagnostics and troubleshooting Diagnostic behavior Symbol Meaning Alarm • Measurement is interrupted. • Signal outputs and totalizers assume the defined alarm condition. A0013961 • A diagnostic message is generated. Warning Measurement is resumed. The signal outputs and totalizers are not affected. A diagnostic message is generated.
  • Page 82: Calling Up Remedial Measures

    Diagnostics and troubleshooting Proline Cubemass C 100 Ethernet/IP 12.3.2 Calling up remedial measures X X X X X X X X X X X X X X 20.50 S801 Supply voltage Menu Diagnostic list Diagnostics 1 S801 Supply voltage Diagnostics 2...
  • Page 83: Diagnostic Information In The Web Browser

    Proline Cubemass C 100 Ethernet/IP Diagnostics and troubleshooting 12.4 Diagnostic information in the Web browser 12.4.1 Diagnostic options Any faults detected by the measuring device are displayed in the Web browser on the home page once the user has logged on.
  • Page 84: Calling Up Remedy Information

    Diagnostics and troubleshooting Proline Cubemass C 100 Ethernet/IP Diagnostic information The fault can be identified using the diagnostic information. The short text helps you by providing information about the fault. Diagnostic information Diagnostic code Status signal Diagnostic number Short text ↓...
  • Page 85: Calling Up Remedy Information

    Proline Cubemass C 100 Ethernet/IP Diagnostics and troubleshooting Xxxxxx/…/…/ Device name: Xxxxxxx Mass flow: kg/h 12.34 Device tag: Xxxxxxx Volume flow: 12.34 m /h ³ Status signal: Function check (C) Xxxxxx Diagnostics 1: C485 Simu... Remedy information: Deactivate... Failure (F)
  • Page 86: Diagnostic Information Via Communication

    Diagnostics and troubleshooting Proline Cubemass C 100 Ethernet/IP 2. On the right in the working area, mouse over the parameter.  A tool tip with remedy information for the diagnostic event appears. 12.6 Diagnostic information via communication interface 12.6.1 Reading out diagnostic information...
  • Page 87 Proline Cubemass C 100 Ethernet/IP Diagnostics and troubleshooting Diagnostic Short text Remedy instructions Status Diagnostic number signal behavior [from the [from the factory] factory] Diagnostic of sensor Sensor temperature 1. Change main electronic Alarm module 2. Change sensor Sensor limit exceeded 1.
  • Page 88 Diagnostics and troubleshooting Proline Cubemass C 100 Ethernet/IP Diagnostic Short text Remedy instructions Status Diagnostic number signal behavior [from the [from the factory] factory] Special event 2 Contact service Alarm Special event 6 Contact service Alarm Special event 10 Contact service...
  • Page 89: Pending Diagnostic Events

    Proline Cubemass C 100 Ethernet/IP Diagnostics and troubleshooting Diagnostic Short text Remedy instructions Status Diagnostic number signal behavior [from the [from the factory] factory] Tubes not oscillating 1. Check electronic Alarm 2. Inspect sensor Medium inhomogeneous 1. Check process cond.
  • Page 90: 12.10 Diagnostic List

    Diagnostics and troubleshooting Proline Cubemass C 100 Ethernet/IP Parameter overview with brief description Parameter Prerequisite Description User interface Factory setting Actual diagnostics 1 diagnostic event has Displays the current diagnostic Symbol for diagnostic – occurred. event along with the behavior, diagnostic diagnostic information.
  • Page 91: 12.11.2 Filtering The Event Logbook

    Proline Cubemass C 100 Ethernet/IP Diagnostics and troubleshooting 12.11.2 Filtering the event logbook Using the Filter options parameter, you can define which category of event messages is displayed in the Events list submenu. Navigation path "Diagnostics" menu → Event logbook → Filter options Filter categories •...
  • Page 92: 12.12 Resetting The Measuring Device

    Diagnostics and troubleshooting Proline Cubemass C 100 Ethernet/IP Info number Info name I1448 Application reference data recorded I1449 Recording application ref. data failed I1450 Monitoring off I1451 Monitoring on I1457 Failed:Measured error verification I1459 Failed: I/O module verification I1460 Failed: Sensor integrity verification...
  • Page 93: 12.13 Device Information

    Proline Cubemass C 100 Ethernet/IP Diagnostics and troubleshooting 12.13 Device information The Device information submenu contains all the parameters that display different information for identifying the device. Navigation "Diagnostics" menu → Device information ‣ Device information Device tag Serial number...
  • Page 94 Diagnostics and troubleshooting Proline Cubemass C 100 Ethernet/IP Parameter Description User interface / User entry Factory setting Extended order code 1 Displays the 1st part of the extended order Character string – code. Extended order code 2 Displays the 2nd part of the extended order Character string code.
  • Page 95: 12.14 Firmware History

    "Manufacturer' s information" document. The manufacturer' s information is available: • In the Download Area of the Endress+Hauser Internet site: www.endress.com → Download • Specify the following details: •...
  • Page 96: Maintenance

    →  112. 13.2 Measuring and test equipment Endress+Hauser offers a wide variety of measuring and test equipment, such as W@M or device tests. Your Endress+Hauser Sales Center can provide detailed information on the services. List of some of the measuring and testing equipment: →  99→  100 13.3...
  • Page 97: Repair

    • The measuring devices have a modular design. • Spare parts are grouped into logical kits with the associated Installation Instructions. • Repairs are carried out by Endress+Hauser Service or by appropriately trained customers. • Certified devices can only be converted to other certified devices by Endress+Hauser Service or at the factory.
  • Page 98: Disposal

    (WEEE), the product is marked with the depicted symbol in order to minimize the disposal of WEEE as unsorted municipal waste. Do not dispose of products bearing this marking as unsorted municipal waste. Instead, return them to Endress+Hauser for disposal under the applicable conditions.
  • Page 99: Accessories

    Various accessories, which can be ordered with the device or subsequently from Endress +Hauser, are available for the device. Detailed information on the order code in question is available from your local Endress+Hauser sales center or on the product page of the Endress+Hauser website: www.endress.com.
  • Page 100: Service-Specific Accessories

    Accessories Description Applicator Software for selecting and sizing Endress+Hauser measuring devices: • Choice of measuring devices for industrial requirements • Calculation of all the necessary data for identifying the optimum flowmeter: e.g. nominal diameter, pressure loss, flow velocity and accuracy.
  • Page 101: Technical Data

    Proline Cubemass C 100 Ethernet/IP Technical data Technical data 16.1 Application The measuring device is intended only for the flow measurement of liquids and gases. Depending on the version ordered, the measuring device can also measure potentially explosive, flammable, poisonous and oxidizing media.
  • Page 102: Input

    Technical data Proline Cubemass C 100 Ethernet/IP 16.3 Input Measured variable Direct measured variables • Mass flow • Density • Temperature Calculated measured variables • Volume flow • Corrected volume flow • Reference density Measuring range Measuring ranges for liquids Measuring range full scale values ...
  • Page 103: Output

    • Operating pressure to increase accuracy (Endress+Hauser recommends the use of a pressure measuring device for absolute pressure, e.g. Cerabar M or Cerabar S) •...
  • Page 104 Technical data Proline Cubemass C 100 Ethernet/IP Web browser Plain text display With information on cause and remedial measures Light emitting diodes (LED) Status information Status indicated by various light emitting diodes The following information is displayed depending on the device version: •...
  • Page 105 Proline Cubemass C 100 Ethernet/IP Technical data Fix Input 5 ms to 10 s (factory setting: 20 ms) Exclusive Owner Multicast Instance Size [byte] Instance configuration: 0x68 O → T configuration: 0x66 T → O configuration: 0x64 Exclusive Owner Multicast...
  • Page 106: Power Supply

    Technical data Proline Cubemass C 100 Ethernet/IP Configurable Input Assembly • Current device diagnostics • Mass flow • Volume flow • Corrected volume flow • Density • Reference density • Temperature • Totalizer 1 • Totalizer 2 • Totalizer 3 ...
  • Page 107: Performance Characteristics

    Proline Cubemass C 100 Ethernet/IP Technical data Power consumption Transmitter Maximum Order code for "Output" Power consumption Option N: EtherNet/IP 3.5 W Current consumption Transmitter Maximum Maximum Order code for "Output" Current consumption switch-on current Option N: EtherNet/IP 145 mA 18 A (<...
  • Page 108 Technical data Proline Cubemass C 100 Ethernet/IP Mass flow (gases) ±0.50 % o.r. Density (liquids) Under reference conditions Standard density calibration Wide-range 2) 3) Density specification [g/cm³] [g/cm³] [g/cm³] ±0.0005 ±0.02 ±0.002 Valid over the entire temperature and density range Valid range for special density calibration: 0 to 2 g/cm³, +5 to +80 °C (+41 to +176 °F)
  • Page 109 Proline Cubemass C 100 Ethernet/IP Technical data Accuracy of outputs The output accuracy must be factored into the measured error if analog outputs are used, but can be ignored for fieldbus outputs (e.g. Modbus RS485, EtherNet/IP). The outputs have the following base accuracy specifications.
  • Page 110 Technical data Proline Cubemass C 100 Ethernet/IP [kg/m ] [°C] [°F] -40 0 40 80 120 160 200 240 280 320 360 400 A0016617 Field density calibration, for example at +20 °C (+68 °F) Special density calibration Temperature ±0.005 · T °C (± 0.005 · (T – 32) °F)
  • Page 111: Installation

    Proline Cubemass C 100 Ethernet/IP Technical data Calculation of the maximum repeatability as a function of the flow rate Flow rate Maximum repeatability in % o.r. ⋅ ½ ZeroPoint ⋅ ³ BaseRepeat A0021340 A0021335 ⋅ ½ ZeroPoint ZeroPoint ⋅ ⋅...
  • Page 112: Process

    Technical data Proline Cubemass C 100 Ethernet/IP Vibration resistance • Oscillation, sinusoidal, following IEC 60068-2-6 • 2 to 8.4 Hz, 3.5 mm peak • 8.4 to 2 000 Hz, 1 g peak • Oscillation, broadband noise following IEC 60068-2-64 • 10 to 200 Hz, 0.003 g •...
  • Page 113 Proline Cubemass C 100 Ethernet/IP Technical data Seals For mounting sets with screwed-on connections: • Viton: –15 to +200 °C (–5 to +392 °F) • EPDM: –40 to +160 °C (–40 to +320 °F) • Silicon: –60 to +200 °C (–76 to +392 °F) •...
  • Page 114: 16.10 Mechanical Construction

    Technical data Proline Cubemass C 100 Ethernet/IP 16.10 Mechanical construction Design, dimensions For the dimensions and installation lengths of the device, see the "Technical Information" document, "Mechanical construction" section. Weight All values (weight exclusive of packaging material) refer to devices with VCO couplings.
  • Page 115 Proline Cubemass C 100 Ethernet/IP Technical data Order code for "Housing", option A "Compact, aluminum, coated" The various cable entries are suitable for hazardous and non-hazardous areas. Cable entry/cable gland Material Cable gland M20 × 1.5 Adapter for cable entry with female thread G ½"...
  • Page 116: 16.11 Human Interface

    Technical data Proline Cubemass C 100 Ethernet/IP Accessories Protective cover Stainless steel, 1.4404 (316L) Safety Barrier Promass 100 Housing: Polyamide Process connections • Fixed flange connections: • EN 1092-1 (DIN 2512N) flange • ASME B16.5 flange • JIS B2220 flange •...
  • Page 117 Proline Cubemass C 100 Ethernet/IP Technical data For some work performed on the measuring device (e.g. electrical connection), it is advisable to disconnect the local display from the main electronics module: 1. Press in the side latches of the local display.
  • Page 118: 16.12 Certificates And Approvals

    The device meets the legal requirements of the applicable EU Directives. These are listed in the corresponding EU Declaration of Conformity along with the standards applied. Endress+Hauser confirms successful testing of the device by affixing to it the CE mark. RCM-tick symbol The measuring system meets the EMC requirements of the "Australian Communications...
  • Page 119: 16.13 Application Packages

    The application packages can be ordered with the device or subsequently from Endress+Hauser. Detailed information on the order code in question is available from your local Endress+Hauser sales center or on the product page of the Endress+Hauser website: www.endress.com.
  • Page 120: 16.14 Accessories

    For an overview of the scope of the associated Technical Documentation, refer to the following: • W@M Device Viewer (www.endress.com/deviceviewer): Enter the serial number from nameplate • Endress+Hauser Operations App: Enter the serial number from the nameplate or scan the 2D matrix code (QR code) on the nameplate Endress+Hauser...
  • Page 121 Proline Cubemass C 100 Ethernet/IP Technical data Standard documentation Brief Operating Instructions Brief Operating Instructions for the sensor Measuring device Documentation code Proline Cubemass C KA01217D Transmitter Brief Operating Instructions Measuring device Documentation code Proline Promass 100 KA01332D Technical Information...
  • Page 122: Index

    Index Proline Cubemass C 100 Ethernet/IP Index Device components ......12 Device description files ..... . . 49 About this document .
  • Page 123 Proline Cubemass C 100 Ethernet/IP Index Web server ......46, 117 Received goods ......13 Electromagnetic compatibility .
  • Page 124 Index Proline Cubemass C 100 Ethernet/IP Menus Power consumption ......107 For measuring device configuration ... . 54 Power supply failure .
  • Page 125 Proline Cubemass C 100 Ethernet/IP Index Simulation ......68 Version ....... . 49 System units .