Fluke 515A Technical Manual page 19

Portable calibrator
Table of Contents

Advertisement

NAVAIR 17–20AQ–70
CALIBRATION CHECKLIST
TEST INST (S)
Fluke 515A Portable Calibrator
NA 17–20AQ–70
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.5
AC Frequency Response Test up to 50 KHz
TI Range
4.5.8
10V
4.5.9
10V
"
10V
4.6
AC Voltage Test
TI Output
Range
4.6.2.2
1V
"
100V
4.7
Frequency Test
TI Output
Volts
4.7.2
10
"
10
"
10
4.8
Resistance Test
TI Output
( )
4.8.11
0
4.8.12
10
"
100
(k )
4.8.12
1
"
(k )
4.8.12
10k
"
100k
(M )
4.8.12
1
"
10
16
MFG.
NOMINAL
(3)
Freq
(DC Volts)
400 Hz
10.000
4 kHz
10.000
50 kHz
10.000
Freq
(%)
400 Hz
0
"
0
Freq
(Hz)
400 Hz
400
4 kHz
4000
50 kHz
50000
(Residual)
––
––
––
––
––
––
––
––
MODEL
MEASURED VALUES
OUT
FIRST RUN
SECOND RUN
TOL (6)
(4)
(5)
(DC Volts)
(%)
(Hz)
( )
( )
( )
( )
SER. NO.
CALIBRATION TOLERANCES
OF
(7)
(DC Volts)
9.996
to
10.004
"
9.990
to
10.010
0.058% or less
0.068% or less
(Hz)
396
to
404
3960
to
4040
47500
to
52500
(m )
150 or less
(6
residual) or less
(60
residual) or less
(m )
(150
residual) or less
( )
(1.5
residual) or less
(15
residual) or less
( )
(150
residual) or less
(7500
residual) or less
Page 2 of 2