NAVAIR 17–20AQ–70
4.4.3 Set the dc voltage standard for a 10V output and adjust it as required to establish a reference null
indication on the 10 V range of the null detector.
4.4.4 For each of the following TI output and voltage divider setting verify that the null detector indication is
within the tolerance limits listed.
4.4.5 Set the equipment control(s) for minimum voltage output, and disconnect the test configuration.

4.5 AC FREQUENCY RESPONSE TEST UP TO 50 KHZ

4.5.1 Connect equipment as shown in Figure 5.
TI
(Perform AC–DC difference corrections as given in the Thermal Transfer Standard's Current
Calibration Report.)
10
TI
Output
Setting
10V x 0.9
"
0.8
"
0.7
"
0.6
"
0.5
"
0.4
"
0.3
"
0.2
"
0.1
2.6
TTS
Figure 5. Frequency Response Test Configuration
Ratio
Setting
0.899999 (X)
0.799999 (X)
0.699999 (X)
0.599999 (X)
0.499999 (X)
0.399999 (X)
0.299999 (X)
0.199999 (X)
0.099999 (X)
AC
DC
NOTE
Tolerance
Limits
270 V or less
240
"
210
"
180
"
150
"
120
"
90
"
60
"
30
"
2.4.3
DC VOLTAGE
CALIBRATOR
SENS
OUT