Fluke SMFT-1000 Calibration Manual page 29

Multifunction pv analyzer
Hide thumbs Also See for SMFT-1000:
Table of Contents

Advertisement

Range
0.5 V to 4 V
5. Verify the calibrator is set to 2-wire, low-resistance output mode.
6. Turn the rotary switch on the DUT to
7. Push
to select Diode mode.
8. Push OPER on the 532xA to set the operate mode, if necessary, and wait for the calibrator
to stabilize.
9. Momentarily push
Wait for the DUT to complete its measurement.
10. Momentarily push
Record the V
DIODE
in the test record. Multiply the I
calibrator (in Ohms) and use this as the reference voltage value to compare V
Ignore the red X or green
11. Set the 532xA calibrator to output the next resistance test point (see
12. Repeat steps
8
13. Set the 532xA calibrator to standby mode.
Table 9. Diode Voltage Test Points
Reference
Resistance [ Ω ]
2.38
4.46
6.52
26.67
on the DUT.
on the DUT to stop continuous measurements made by the DUT.
on the DUT and the maximum current value indication on the 532xA (I
value (in Amps) by the resistance output by the 532xA
D
. They do not indicate a pass or fail for this calibration.
through
11
for all test points.
Test Points
[V]
0.5
0.38
1.00
0.85
2.00
1.80
5.60
5.22
.
Note
Multifunction PV Analyzer
Performance Tests
Display Readings [V]
Lower
Upper
0.63
1.15
2.20
5.98
DIODE
Table
9).
)
D
.
25