Fluke SMFT-1000 Calibration Manual page 11

Multifunction pv analyzer
Hide thumbs Also See for SMFT-1000:
Table of Contents

Advertisement

8. Set the 532xA calibrator to output the first resistance test point (see
2-wire, ground bond resistance (Z GND) output mode.
Range
0 Ω to 19.99 Ω
20 Ω to 199.0 Ω
200 Ω to 2000 Ω
9. Push
on the DUT to set the One Shot mode.
10. Push OPER on the 532xA to set the operate mode, if necessary, and wait for the calibrator
to stabilize.
Figure 2. R
Connections
LO
Fluke 5320A
LO
L
,
Fluke SMFT-1000
Table 2. R
Test Points
LO
Test Points
[ Ω ]
0.35
0.96
1.70
17.00
90.0
170.0
470
900
1700
Multifunction PV Analyzer
Performance Tests
L
,
METER
Table
Display Reading [ Ω ]
Lower Limit
0.32
0.38
0.92
1.00
1.65
1.75
16.64
17.36
88.0
92.0
166.4
173.6
445
496
853
947
1613
1787
+/-
2) and set to
Upper Limit
7