Fluke SMFT-1000 Calibration Manual page 32

Multifunction pv analyzer
Hide thumbs Also See for SMFT-1000:
Table of Contents

Advertisement

SMFT-1000
Calibration Manual
5. Push
until Voltage shows in highlight on the left-side menu.
6. Set the 552xA calibrator to output the first voltage test point (see
Range
0 V to 99.9 V
100 V to 1000 V
5 V to 99.9 V
100 V to 700 V
5 V to 99.9 V
100 V to 700 V
7. Push OPER on the 552xA to set the operate mode, if necessary, and wait for the calibrator
to stabilize.
28
Table 11. AC/DC Voltage Test Points
Frequency [Hz]
5.0
50.0
75.0
97.0
0
105
120
400
970
5.0
50.0
75.0
97.0
50
105
250
400
680
5.0
50.0
75.0
97.0
60
105
250
400
680
Test Points
[V]
Lower
4.7
48.6
72.9
94.4
100
115
388
944
4.7
48.6
72.9
94.4
100
242
388
661
4.7
48.6
72.9
94.4
100
242
388
661
Table
11).
Display Readings [V]
Upper
5.3
51.5
77.1
99.6
110
125
412
996
5.3
51.5
77.1
99.6
110
258
412
699
5.3
51.5
77.1
99.6
110
258
412
699