Fluke SMFT-1000 Calibration Manual page 17

Multifunction pv analyzer
Hide thumbs Also See for SMFT-1000:
Table of Contents

Advertisement

Range
0.2 M Ω to 99.99 M Ω
100 M Ω to 199.9 M Ω
0.2 M Ω to 99.99 M Ω
100 M Ω to 199.9 M Ω
0.2 M Ω to 99.99 M Ω
100 M Ω to 199.9 M Ω
Table 5. R
Method I (No Voltage) Test Points
INS
Test Voltage
[V]
50
100
250
Test Points
[M Ω ]
0.25
0.19
0.40
0.33
1.60
1.47
5.00
4.70
50.00
47.45
95.00
90.20
112.0
100.3
160.0
143.5
180.0
161.5
0.25
0.19
0.80
0.71
2.00
1.85
10.00
9.45
50.00
47.45
95.00
90.20
112.0
100.3
150.0
134.5
180.0
161.5
0.25
0.19
20.00
18.95
50.00
47.45
95.00
90.20
112.0
100.3
160.0
143.5
180.0
161.5
Multifunction PV Analyzer
Performance Tests
Display Readings [M Ω ]
Lower
Upper
0.31
0.47
1.73
5.30
52.55
99.80
123.7
176.5
198.5
0.31
0.89
2.15
10.55
52.55
99.80
123.7
165.5
198.5
0.31
21.05
52.55
99.80
123.7
176.5
198.5
13

Advertisement

Table of Contents
loading

Table of Contents