Fluke SMFT-1000 Calibration Manual page 21

Multifunction pv analyzer
Hide thumbs Also See for SMFT-1000:
Table of Contents

Advertisement

Range
0.2 M Ω to 99.99 M Ω
100 M Ω to 199.9 M Ω
200 M Ω to 999 M Ω
7. Verify the calibrator is set to high resistance output mode.
8. Set the 2290-5 to output the first voltage test point (see
The 2290-5 output must be within 5 % of nominal.
9.
Push OPER on the 532xA and 2290-5 to set the operate mode, if necessary, and wait for
the calibrator to stabilize.
10. Push and hold
Wait for the DUT to finish both the + and - measurements. Record only the plus indication
on the DUT in the test record.
Ignore the red X or green
11. Set:
532xA calibrator to output the next resistance test point
2290-5 to output the next voltage setting
voltage output of the DUT to the next voltage setting
See test points in
The 2290-5 output must be within 5 % of nominal.
12. Repeat steps
9
13. Set the 532xA and 2290-5 to standby mode.
Table 6. R
Method I (With Voltage) Test Points
INS
DUT Voltage
[V]
Voltage [V]
40
50
500
40
500
500
40
1000
500
on the DUT until it beeps (~1 second).
. They do not indicate a pass or fail for this calibration.
Table
6.
through
11
for all test points.
Power
Test Points
Supply
[M Ω ]
50.00
150.0
500
Note
Multifunction PV Analyzer
Performance Tests
Display Readings [M Ω ]
Lower
47.45
47.45
134.5
134.5
395
395
Table
6).
Upper
52.55
52.55
165.5
165.5
605
605
17