Range
0.2 M Ω to 99.99 M Ω
100 M Ω to 199.9 M Ω
0.2 M Ω to 99.99 M Ω
100 M Ω to 199.9 M Ω
Table 7. R
Method II (No Current) Test Points
INS
Test Voltage
[V]
50
100
Test Points
[M Ω ]
0.25
0.19
0.40
0.33
1.00
0.90
5.00
4.70
10.00
9.45
25.00
23.70
50.00
47.45
70.00
66.45
95.00
90.20
112.0
100.3
160.0
143.5
180.0
161.5
0.25
0.19
0.80
0.71
12.00
11.35
50.00
47.45
80.00
75.95
95.00
90.20
112.0
100.0
160.0
143.5
180.0
161.5
Multifunction PV Analyzer
Performance Tests
Display Readings [M Ω ]
Lower
Upper
0.31
0.47
1.10
5.30
10.55
26.30
52.55
73.55
99.80
123.7
176.5
198.5
0.31
0.89
12.65
52.55
84.05
99.80
123.7
176.5
198.5