Fluke SMFT-1000 Calibration Manual page 19

Multifunction pv analyzer
Hide thumbs Also See for SMFT-1000:
Table of Contents

Advertisement

Range
Voltage Output
Accuracy
8. Verify the calibrator is set to high resistance output mode.
9. Push OPER on the 532xA to set the operate mode, if necessary, and wait for the calibrator
to stabilize.
10. Push and hold
Wait for the DUT to finish both the + and - measurements. Record both the + and
- indication on the DUT in the test record. Also record the maximum voltage measurement
indication on the 532xA calibrator in the test record.
Ignore the red X or green
11. Set the 532xA calibrator to output the next resistance test point and, if applicable, set the
voltage output of the DUT to the next voltage setting (see
12. Repeat steps
9
13. Set the 532xA calibrator to standby mode.
Table 5. R
Method I (No Voltage) Test Points (cont.)
INS
Test Voltage
[V]
50
100
250
500
1000
on the DUT until it beeps (~1 second).
. They do not indicate a pass or fail for this calibration.
through
11
for all test points.
Test Points
[M Ω ]
0.25
50.0
180
50.0
0.25
100.0
180
100.0
0.25
250.0
180
250.0
0.50
500.0
830
500.0
1.00
1000
830
1000
Note
Table
Multifunction PV Analyzer
Performance Tests
Display Readings [V]
Lower
Upper
62.5
62.5
125.5
125.5
312.5
312.5
625.0
625.0
1250.0
1250.0
5).
15