Section 6 - Specifications
6.1

General

Temperature Range:
Accuracy @ 22°C (72°F)

Optical Field of View:

Repeatability:
Spectral Response:
Response Time:
Emissivity Range:
Operating Ambient Temperature
Operating Relative Humidity:
Water Flow Rate for OS100-WC: 0.25 GPM, room temperature
Thermal Shock:
Warm Up Period:
Air Flow Rate for OS100-AP:
Power:
Analog Outputs
Output Load requirements
Transmitter Housing:
Sensor Head Housing:
Alarm Output:
Alarm Set Point:
Alarm Deadband:
Alarm Indication:
ambient temperature &
emissivity of 0.95 or greater:
Main Transmitter:
Sensor Head:
Sensor Head (-HT Model):
Sensor Head with OS100-WC
(Water Cooling Jacket):
MV:
K:
MA:
V1:
V2:
Min. Load (0 to 5Vdc):
Max. Load (4 to 20 mA):
OS100E Series - Specifications
-18 to 538°C (0 to 1000°F)
± 2% of Rdg. or 2.2°C (4°F) whichever is
greater
6:1 (Distance/Spot Size)
± 1% of Rdg.
5 to 14 microns
100 msec (0 to 63% of final value)
0.1 to 1.00, adjustable
0 to 50°C (32 to 122°F)
0 to 70°C (32 to 158°F)
0 to 85°C (32 to 185°F)
0 to 200°C (32 to 392°F)
Less than 95% RH, non-condensing
About 30 minutes for 25°C
abrupt ambient temperature change
3 minutes
1 CFM (0.5 Liters/sec.)
12 to 24 Vdc @ 150 m
(0-10 Vdc output requires 14 Vdc)
1 mV/°F or 1 mV/°C
K Type TC
4 to 20 mA
0 to 5 Vdc
0-10 Vdc
1 K-Ohms
(Supply Power - 4 )/20 mA
NEMA-4 & IP65, Die Cast Aluminum
NEMA-4; Anodized Aluminum
Open Drain, 100 mA
0 to 100%, Adjustable via P3
12°C (10°F)
Display Flashing
6
6-1