Post-Installation Control; Adjusting The Cephalostat Head; Adjusting The Inclination Of The Cephalostat Head - Kodak 9000C Installation Manual

Extraoral imaging system
Table of Contents

Advertisement

Post-Installation Control

5–10
INSTALLING THE CEPHALOSTAT
Post-Installation Control
Before the post-installation control, check that:
The installation of the unit is complete.
The installation of the KODAK dental imaging software is complete.
You have the test tool.
The unit is on.
WARNING
You MUST correctly calibrate the rotative arm of KODAK 9000
or KODAK 9000 3D. If the arm calibration is not correctly
executed the cephalometric mode will not be operational. (see
KODAK 9000 and KODAK 9000 3D Service Guide (SM712) Ed02,
Using the Technical Menu, Calibration procedures, "Arm
Calibration")

Adjusting the Cephalostat Head

Adjusting the Inclination of the Cephalostat Head

To adjust the cephalostat head inclination, follow these steps:
1. Position the head clamps in a frontal AP position.
2. Access the Cephalometric Acquisition Window. In the Program pane:
Click
.
Click
for a 30 x 30 format.

Advertisement

Table of Contents
loading

This manual is also suitable for:

9000c 3d

Table of Contents