Electrical Requirements - Kodak 9000C Installation Manual

Extraoral imaging system
Table of Contents

Advertisement

Electrical Requirements

WARNING*
You MUST select the operating voltage when placing an order. The
operating voltage CANNOT be modified on site.
The unit can operate at:
100/110/130 V 50/60 Hz
230/240 V 50/60 Hz
Table 4–1
Optional Operating Voltages of the Unit
Nominal voltage *
(no load)
100 V - 130 V
230 V - 240 V
CAUTION
The power supply line must be equipped with a connection box that ensures a
constant connection. It must not be possible to connect the unit to the power
supply without using a tool. The unit must be protected against any
accidental disconnection.
If other units are installed on the same line, interference and voltage fluctuations can cause the
radiological unit to operate abnormally. We strongly recommend that a separate electrical line
be dedicated to supply power to the KODAK 9000 and KODAK 9000 3D unit.
This line should be protected by a circuit breaker with a maximum current of:
16 A at 230/240 V
20 A at 110/130 V
A differential circuit breaker of 30 mA
Minimum
90 V
207 V
KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System Installation Guide (SM714)_Ed 01
Maximum line
Maximum
current
143 V
264 V
Electrical Requirements
20 A
10 A
4–3