Endress+Hauser MemoLink Operating Instructions Manual

Endress+Hauser MemoLink Operating Instructions Manual

Sensor terminal box for memobase plus cyz71d

Advertisement

Quick Links

BA01134C/07/EN/03.20
71507627
2020-11-16
0
Products
Operating Instructions
MemoLink
Sensor terminal box for Memobase Plus CYZ71D
Solutions
Services

Advertisement

Table of Contents
loading

Summary of Contents for Endress+Hauser MemoLink

  • Page 1 Products Solutions Services BA01134C/07/EN/03.20 71507627 2020-11-16 Operating Instructions MemoLink Sensor terminal box for Memobase Plus CYZ71D...
  • Page 3: Table Of Contents

    Electrical connection ..10 Connecting MemoLink ....10 Repair ......11 Return .
  • Page 4: About This Document

    About this document MemoLink About this document Warnings Structure of information Meaning This symbol alerts you to a dangerous situation. DANGER Failure to avoid the dangerous situation will result in a fatal or serious injury. Causes (/consequences) If necessary, Consequences of non- compliance (if applicable) ‣...
  • Page 5: Basic Safety Instructions

    Memobase Plus cannot be used to replace a process transmitter, as communication with the control system is not supported. MemoLink acts as the Memosens/USB interface converter for the Memobase Plus software. MemoLink supports Memosens sensors with Ex approval and without Ex approval.
  • Page 6: Product Safety

    2.5.2 Electrical equipment in hazardous areas • The MemoLink must never be used in hazardous areas! The MemoLink is not designed as a field device for installation in industrial environments. • The MemoLink, which is approved in accordance with EC type-examination certificate BVS 12 ATEX 079 X, ensures that the "intrinsic safety"...
  • Page 7: Product Description

    Product description Product design 1 to 4 MemoLink sensor terminal boxes, with the same number of Memosens sensors (1 to 4), can be connected per license. The following components are required in each case: • Computer with the Memobase Plus software installed (not supplied) •...
  • Page 8: Incoming Acceptance And Product Identification

    Incoming acceptance and product identification MemoLink Incoming acceptance and product identification Incoming acceptance Verify that the packaging is undamaged.  Notify the supplier of any damage to the packaging. Keep the damaged packaging until the issue has been resolved. Verify that the contents are undamaged.
  • Page 9: Scope Of Delivery

    • Quick installation guide • Operating Instructions for MemoLink • Depending on the version ordered: • License • 0 to 4 MemoLink sensor terminal boxes • 0 to 4 Memosens cable CYK20 • 0 to 4 USB cables ‣ If you have any queries: Please contact your supplier or local sales center.
  • Page 10: Installation

    Power / Data A0025837  2 Dimensions of MemoLink in mm (in) The MemoLink sensor terminal boxes can be stacked on top of one another. In such situations, the "Power / Data" LED is still easily visible. Electrical connection Connecting MemoLink Connect the mini USB connector to the mini USB jack on the MemoLink.
  • Page 11: Repair

    MemoLink Repair Connect the M12 connector to the M12 jack on the MemoLink. Use the flexible Memosens laboratory cable CYK20 or the Memosens process cable CYK10.  A0031653 Cable with mini USB connector Cable with M12 connector Connect the USB connector to the USB jack on your computer.
  • Page 12: Technical Data

    Max. 30 m (98 ft) Power supply 8.4.1 Supply voltage The PC supplies power to the sensor(s) and MemoLink(s) via the USB cable and enables bidirectional transfer of Memosens data. An active USB hub must have a power unit. 8.4.2 Power supply •...
  • Page 13: Performance Characteristics

    For detailed information on "Measured error", see the documentation for the connected sensor. MemoLink only transmits data digitally so no measured data can be corrupted. The measuring signal is converted to digital data in the sensor, which means that the measured values are not affected by MemoLink, the cable or the software.
  • Page 16 *71507627* 71507627 www.addresses.endress.com...