Setting Up The Stable Clock; Setting Up A Pxi-6608 Clock Source - National Instruments NI 6601 Manual

Calibration procedure
Hide thumbs Also See for NI 6601:
Table of Contents

Advertisement

Setting Up the Stable Clock

© National Instruments Corporation
The frequency measurement will have an uncertainty that is equal to the
sum of the stable clock error and the quantization error caused by the
counter. The clock error should be 1 ppm or less, as specified in the
Equipment
section of this document. The quantization error, which is the
inherent uncertainty in digitizing an analog value, will always be up to
50 ns. Using a measurement duration of 50 ms gives a quantization error
as follows:
The sum of the stable clock error, 1 ppm, and the quantization error,
1 ppm, gives a total measurement uncertainty of 2 ppm. You can reduce
measurement uncertainty by using a more stable clock, by increasing the
measurement duration, or both. The calculations presented at the end of this
document allow for a 2 ppm measurement uncertainty.
Both Traditional NI-DAQ and LabVIEW programming instructions are
presented in the following sections. Choose the appropriate instructions for
your programming environment.
This section describes how to set up the clock. NI recommends using the
PXI-6608 to provide the stable clock. You can use an alternative stable
clock if it meets the specifications described in the
of this document. Instructions for both clock alternatives are included in the
following two sections, Setting Up a PXI-6608 Clock Source and
Up an Alternative Clock

Setting Up a PXI-6608 Clock Source

To set up a PXI-6608 as the stable clock source, complete the following
steps:
1.
Power on the 6608 device.
2.
Wait 15 minutes to allow the measurement circuitry in the device
to warm up.
50 ns
1 ppm
------------- -
=
50 ms
Source.
5
Test Equipment
section
Setting
NI 6601/6602 Calibration Procedure
Test