Connecting An Alternative Clock Source To The Ni 6601/6602 Device; Measuring The Frequency Of The Ni 6601/6602 Device - National Instruments NI 6601 Manual

Calibration procedure
Hide thumbs Also See for NI 6601:
Table of Contents

Advertisement

Measuring the Frequency of the NI 6601/6602 Device

If you are verifying the PXI-6602 on a PXI chassis, the measured frequency is that
Note
of the PXI backplane clock instead of the onboard oscillator. To verify the onboard
oscillator, you must calibrate the PXI-6602 on a PCI chassis using a CompactPCI-to-PCI
adapter.
Traditional NI-DAQ Function Call
Call
GPCTR_Set_Application
following parameters:
deviceNumber: The value set by MAX
gpctrNum:
ND_COUNTER_0
application:
ND_SIMPLE_EVENT_CNT
© National Instruments Corporation
Connecting an Alternative Clock Source to the
NI 6601/6602 Device
To connect an alternative clock source to the NI 6601/6602 device,
complete the following steps:
1.
Connect the NI 6601/6602 device to a 68-pin breakout box using the
SH68-68-D1 cable.
2.
Connect the clock source to pins 7 and 41 of the 68-pin breakout box
using twisted-pair wires.
Connect the clock to pin 7. Pin 7 corresponds to PFI_35, which is
the SOURCE of counter 1 on all NI 6601/6602 devices.
Connect the ground of the clock to pin 41, which is digital ground.
To measure the frequency of the NI 6601/6602 device for verification,
complete the following steps:
1.
Configure counter 0 to perform simple event counting.
with the
LabVIEW Block Diagram
9
NI 6601/6602 Calibration Procedure