Comparing The Measured Frequency To The Ni 6601/6602 Device Specifications - National Instruments NI 6601 Manual

Calibration procedure
Hide thumbs Also See for NI 6601:
Table of Contents

Advertisement

Traditional NI-DAQ Function Call
Call
GPCTR_Control
parameters:
deviceNumber: The value set by MAX
gpctrNum:
ND_COUNTER_1
action:
ND_RESET
Comparing the Measured Frequency to the NI 6601/6602 Device
Specifications
CVI, National Instruments, NI, ni.com, and LabVIEW are trademarks of National Instruments
Corporation. Refer to the Terms of Use section on ni.com/legal for more information about
National Instruments trademarks. The mark LabWindows is used under a license from Microsoft
Corporation. Windows is a registered trademark of Microsoft Corporation in the United States and
other countries. Other product and company names listed are trademarks or trade names of their
respective companies. For patents covering National Instruments products/technology, refer to the
appropriate location: Help»Patents in your software, the patents.txt file on your media, or the
National Instruments Patent Notice at ni.com/patents.
© 2001–2009 National Instruments Corporation. All rights reserved.
14. Reset counter 1 on the PXI-6608 to its original state, if necessary.
with the following
You have now completed measuring the frequency of the crystal oscillator
on the NI 6601/6602 device.
Compare the calculated frequency from step 11 in the previous section with
the following values:
If the frequency you measure falls between the upper limit of
20,001,500 Hz and the lower limit of 19,998,500 Hz, the device is
operating within its specifications.
If the frequency you measure is above 20,001,500 Hz, the device is
nonfunctional. Return the device to NI for repair or replacement.
If the frequency you measure is below 19,998,500 Hz, the device is
nonfunctional. Return the device to NI for repair or replacement.
You have completed verifying the operation of the NI 6601/6602 device.
LabVIEW Block Diagram
"TASK ID" in connects to "TASK ID" out of step 3e
in the
Setting Up the Stable Clock
section.
370270C-01
Oct09