Service Guide

Setup Tab

1
2
3
4
5
6
The Setup tab is used to enable or disable certain applications. It can be accessed by both the Key Operator and the
Full Service Operator.
#
Component
1
Scan Mode
2
Image options
138
7
Set Scan Mode to Scan for normal scanning of cassettes. Select Demo to simulate a
scan that displays example images. No cassette is required in Demo mode. The images
are located in C:\Program Files\Kodak\KodakQC\CR 360\Demo Images folder.
Sets image processing options for the scanned image.
KODAK Image Processing is a mandatory image enhancement tool which includes six
different looks. Looks are determined in the Anatomical tab.
Black Surround activates automatic detection of the collimation regions and surrounds
them with black borders. Default: checked.
Multi Exposure on a single screen displays all sub-image fields and black-masks the
collimation. Default: unchecked.
Grid Suppression applies an algorithm to remove grids lines caused by using stationary
grid. Grid Suppression algorithm requires 103 lp/mm. Default: checked.
16-bit linear data saves the raw unprocessed images as a 16 bit image in C:\Program
Files\Kodak\KodakQC\Images\Linear Images. Service use only.
Framing removes the white screen overscan. Default: checked.
OCPC process corrects for collector profile (uniformity). It can be activated if Framing is
activated. Default: checked.
8
9
10
11
12
Description
6H4869