Service Guide
Section 6: Preventive Maintenance
Cleaning the Rollers
Roller Cleaning Cycles
The roller-cleaning device enables you to clean the rollers that feed the storage phosphor screen from the cassette
into the scanner. Clean the rollers periodically to remove dust and small particles.
Every 1,000 scanning cycles the following message appears.
If you do not clean the rollers, the message appears every 100 scanning cycles. Cleaning the rollers resets the roller
cleaning counter.
Important
Not cleaning the rollers periodically will reduce the image quality.
Roller Cleaning Procedure
Note
Perform the roller cleaning procedure three times.
The roller-cleaning device includes the following items:
• Cleaning tray
• Cleaning plate with adhesive strips inside a protective envelope
[1] Remove any cassettes and screens from the scanner.
[2] Start the KODAK QC software.
[3] Go to Settings>User.
130