Electrical Hazards; Explosion And Implosion Hazards; Overheating; Laser Safety Instructions - Kodak CR 360 Service Manual

Table of Contents

Advertisement

Service Guide

Electrical Hazards

Caution
• Do not remove or open system covers or plugs. Internal circuits use high voltage capable of causing serious
injury.
• Fuses blown within 36 hours of being replaced by a qualified technician may indicate malfunctioning electrical
circuits within the system. Have the system checked by qualified service personnel. Do not attempt to replace
any fuse.
• Fluids that seep into the active circuit components of the system may cause short circuits that can result in
electrical fires. Therefore, do not place any liquid or food on any part of the system.

Explosion and Implosion Hazards

Caution
Caution
• Do not operate the equipment in the presence of explosive liquids, vapors, or gases.
• Do not plug in or turn on the system if hazardous substances are detected in the environment. If these
substances are detected after the system has been turned on, do not attempt to turn off the unit or unplug it.
Evacuate and ventilate the area before turning off the system.

Overheating

Do not block the air circulation around the scanner. Always maintain at least 6 inches (15 cm) clearance around the
scanner to prevent overheating and damage to the system.

Laser Safety Instructions

• During nominal operation, the unit is closed and sealed with a protective cover for safety reasons.
• During nominal operation, the cover should not be removed. Removing of the cover shall be done only for service
purposes, and by a qualified technician for service operations.
• Service operations that do not require the laser should be done without activating the laser unit. Switching OFF
the DIP switch on the laser board will disconnect the power supply to the laser, and deactivate the laser unit for
service operations.
• In case the laser must be operated during service operation, the service technician shall make sure that the
optical unit is located within the scanner drum, where the laser beam is blocked.
Laser Warning
When a service operation is taking place with the cover removed, disconnect the laser according to the procedure in
Activating the Scanner Without the Cover (Laser Deactivated) on Page
service procedure, wear protective safety glasses at all times. The required laser safety eyewear must be intended
for HeliumNeon/PDT lasers, have an optical density of 4-5 wavelengths of 610-695 nm, and be marked as having
CE approval.
10
26. If the laser must be activated during the
6H4869

Advertisement

Table of Contents
loading

Table of Contents