GE Responder 2000 Service Manual page 82

Defibrilator/monitor
Hide thumbs Also See for Responder 2000:
Table of Contents

Advertisement

Test Description
Patient leakage current measurements: SpO
TP1 connected to "SpO
" (all SpO
2
Polarity: Normal
Neutral: Normal (Closed)
Ground: Normal (Closed)
Polarity: Normal
Neutral: Open
Ground: Normal (Closed)
Polarity: Normal
Neutral: Normal (Closed)
Ground: Open
Polarity: Reverse
Neutral: Normal (Closed)
Ground: Open
Polarity: Reverse
Neutral: Open
Ground: Normal (Closed)
Polarity: Reverse
Neutral: Normal (Closed)
Ground: Normal (Closed)
2025653-048 Revision B
Section 5: Performance Verification and Safety Testing
S1: DPDT
S2: SPDT
Polarity
Neutral
a) Normal
a) Closed
b) Reverse
b) Open
Position
Position
a
b
a
b
2
connections tied together), Type BF defibrillation proof limits
2
X
X
X
X
X
X
X
X
X
X
X
X
Responder
2000
S3: SPDT
S4: SPDT
Ground
Test point
a) Closed
a) Internal
b) Open
b) TP1
Position
Position
a
b
a
b
X
X
X
X
X
X
X
X
X
X
X
X
Maximum
Limits
(uA)
EN/IEC
UL
100
100
500
500
500
500
500
500
500
500
100
100
Page 82

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents