Enclosure (Chassis) leakage
Perform the following tests, with the applicable switch settings on the "Safety Tester" and record the results in the
applicable test form(s).
UUT power switch in the "ON" position.
TP1 connected to accessible metal parts (see tables below for individual test points).
Test Description
Enclosure (Chassis) leakage current measurements:
TP1 connected to "RS-232 DB9 shell and Metal part of Printer.
Polarity: Normal
Neutral: Normal (Closed)
Ground: Normal (Closed)
Polarity: Normal
Neutral: Open
Ground: Normal (Closed)
Polarity: Normal
Neutral: Normal (Closed)
Ground: Open
Polarity: Reverse
Neutral: Normal (Closed)
Ground: Open
Polarity: Reverse
Neutral: Open
Ground: Normal (Closed)
Polarity: Reverse
Neutral: Normal (Closed)
Ground: Normal (Closed)
2025653-048 Revision B
Section 5: Performance Verification and Safety Testing
S1: DPDT
S2: SPDT
Polarity
Neutral
a) Normal
a) Closed
b) Reverse
b) Open
Position
Position
a
b
a
b
X
X
X
X
X
X
X
X
X
X
X
X
Responder
2000
S3: SPDT
S4: SPDT
Ground
Test point
a) Closed
a) Internal
b) Open
b) TP1
Position
Position
a
b
a
b
X
X
X
X
X
X
X
X
X
X
X
X
Maximum
Limits
(uA)
EN/IE
UL
C
100
100
500
300
500
300
500
300
500
300
100
100