GE Responder 2000 Service Manual page 74

Defibrilator/monitor
Hide thumbs Also See for Responder 2000:
Table of Contents

Advertisement

Leads
3-7
Patient Auxiliary Current DC (OS)
3-7
Patient Auxiliary Current DC (NC)
3-7
Patient Auxiliary Current DC OE
3-7
Patient Auxiliary Current DC OSRM
3-7
Patient Auxiliary Current DC NCRM
3-7
Patient Auxiliary Current DC OERM
1-2
Patient Leakage Current DC (OS)
1-2
Patient Leakage Current DC NC
1-2
Patient Leakage Current DC OE
1-2
Patient Leakage Current DC OSRM
1-2
Patient Leakage Current DC NCRM
1-2
Patient Leakage Current DC OERM
3-7
Patient Leakage Current DC (OS)
3-7
Patient Leakage Current DC (NC)
3-7
Patient Leakage Current DC OE
3-7
Patient Leakage Current DC OSRM
3-7
Patient Leakage Current DC NCRM
3-7
Patient Leakage Current DC OERM
8
Patient Leakage Current DC (OS)
8
Patient Leakage Current DC (NC)
8
Patient Leakage Current DC OE
8
Patient Leakage Current DC OSRM
8
Patient Leakage Current DC NCRM
8
Patient Leakage Current DC OERM
2.
Plug Cord from 120V, 60hz receptacle of Safety Analyzer into the wall outlet.
3.
Plug cord from Auxiliary Power to Equipment Under Test receptacle of Safety Analyzer into Step-up
transformer output (264VAC).
4.
Plug power cord from Responder 2000 into outlet on front of Safety Analyzer.
5.
Plug jumper cable from ENCL connector to EARTH connector.
6.
Plug the 5 ECG cable leads from ECG Cable Port into Patient Lead connectors 3-7.
7.
Plug cable from Dfib Paddle Port into Patient Lead connectors 1 and 2.
8.
(For SpO
equipped Responder 2000 only) Plug cable from Sp02 connector to Patient Lead connector 8.
2
2025653-048 Revision B
Section 5: Performance Verification and Safety Testing
Name
Responder
2000
Limit
50 A
10 A
50 A
50 A
10 A
50 A
50 A
10 A
50 A
50 A
10 A
50 A
50 A
10 A
50 A
50 A
10 A
50 A
50 A
10 A
50 A
50 A
10 A
50 A
Page 74

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents