SECTION 4: Repair
Assembly Step
Details
Snap the two black
Brackets into place on the
Lower Frame.
Note: Check the Power
Supply green ground wire
to ensure it is properly
soldered on both ends and
is in good condition.
Place the Power Supply on
the Brackets.
Note: The Power Supply is
not secured until the Base
Figure 24: Brackets Installed
Figure 25: Power Supply Installed
Cover is installed.
Place the AC Cable
Connector into the Lower
Frame as shown.
Note: The single ground pin
must be at the top.
Figure 26: AC Cable Connector
Figure 27: Proper Connector
Orientation
Insert the Power Button
and Control Button
assemblies into the Lower
Frame.
Figure 28: Power and Control Button Assemblies Installed
Pass Therapy Power
Cable into the Lower
Frame as shown. Pull the
slack through, leaving
enough length to connect to
Power Control Board
(installed in the next step).
Figure 29: Therapy Power Cable
2025653-048 Revision B
Responder
2000
Page 37