SECTION 4: Repair
Assembly Step
Details
Place the Main Board and
Printer into the Front
Body.
Tuck the ribbon cables
connected to J210 and
J218 as shown.
Secure the Main Board
with seven Philips screws
Figure 12: Cable
and install the copper
Detail
Figure 11: Main Board
Contact Spring as shown.
Figure 13: Copper
Installed
Contact Spring
Note: Ensure the printer
Installation Detail
ribbon cable does not twist
during installation.
Turn the assembly over
(support the Printer so the
cable does not twist) and
secure the printer with three
Phillips screws.
Note: Tighten the center
screw first to properly align
the Printer.
Figure 14: Printer Screws
Connect the speaker cable
to Main Board J216.