Detection Lead:
Lead Selection:
Heart Rate Range:
Display Frequency Response (ECG Leads):
Display Frequency Response (paddles):
ECG Effective Sampling Rate:
Post Defibrillation Recovery:
Heart Rate Alarms:
Leads Off Sensing
ECG Leads and paddles leakage current
Dynamic Range: Input ECG signal amplitude:
Dynamic Range: DC Offset voltage:
Asystole Threshold
Rechargeable Battery
Battery Voltage:
Chemistry:
Compatibility:
Battery Capacity:
Battery Charge Time:
Battery Standby:
Battery Life:
Battery Weight:
AC Power Supply
Universal input:
Printer
Speed
Paper Size
2025653-048 Revision B
SECTION 8: Specifications and Safety
Through either 3-lead and 5-lead patient cable, pads, or
paddles
3-Lead Cable: (I, II, III) and 5-Lead (I, II, III, aVL, aVR,
aVF, and V), PADS: (Modified Lead II)
25 to 300 bpm
0.5 to 100 Hz (± 10 %)
3 Hz to 33 Hz (± 10 %); (-3db)
1000 samples per second (ECG Leads), 128 samples per
second (paddles)
8 seconds
Off, 5-300 bpm
DC current; Sensing Leads <0.1 uA; Reference Lead
<1uA; when all leads are connected
Less than 10uA in normal; Less than 50uA in single fault
condition
± 5mV
± 500 mV ECG from ECG cable; ± 1000 mV (ECG from
pads, paddles and spoons
0.2mV (±0.1mV)
11.1 V Nominal
Lithium-ion
Compatible with Responder 2000
50 shocks, or 240 minutes monitoring time, or 72 minutes
monitoring time with pacing
8 hours in Responder 2000, 4 hours in external charger,
20 hours for calibration cycle in external charger
6 months
2.5 years or 300 Battery charge-discharge cycles,
whichever comes first
1 lb. 3 oz; .54 kg
100V to 240V~
50Hz-60Hz
200VA
25 mm/S feed rate
60 mm paper width
Responder
2000
Page 101