Although the device zero is very stable from the factory the user has the ability to remove any
residual zero offset that may exist by performing the Zero point adjustment.
C.1.11
Changing parameter settings using SIMATIC PDM
SIMATIC PDM monitors the process values, alarms and status signals of the device. It allows you
to display, compare, adjust, verify, and simulate process device data; also to set schedules for
calibration and maintenance.
Parameters in SIMATIC PDM are identified by name and organized into function groups, similar
to the structure of the local display (HMI).
Within SIMATIC PDM, parameters can be found in a structured view (which can be accessed
when the device is offline), or in the PDM menus (when device is online): Device, View,
Diagnostics.
See:
• SIMATIC PDM structure view image (Page 112)
• Parameters accessed via PDM menus (Page 112)
Note
• Clicking button "Cancel" during an upload from device to SIMATIC PDM will result in some
parameters being updated.
• While the device is in Edit view the output remains active and continues to respond to
changes in the device.
1. Launch SIMATIC PDM, connect to device, and upload data from device.
2. Adjust parameter values in the parameter value field then press "Enter" key. The status fields
read "Changed".
3. Open the Device menu, click "Download to device...". When complete, use "File > Save" to
save settings offline. The status fields are cleared.
SITRANS FST020 IP65 (NEMA 4X)
Operating Instructions, 05/2022, A5E41425845-AC
C.1 Commissioning with SIMATIC PDM
SIMATIC PDM
111