Chapter 4 - Troubleshooting and Maintenance
Clean the analyzer at least every week and the docking station (including accessories) at
least every month.
Always allow sufficient incubation time for decontamination.
Cleaning the analyzer outer surfaces
INSTRUCTIONS
1. Clean the analyzer outer surfaces with a cloth or wipe, dampened with one of the
following:
‒ 70% isopropanol alcohol
‒ 70% ethanol alcohol
‒ Antibacterial soap and water
Do not allow liquid to drip into the analyzer.
2. Allow enough time after cleaning for the equipment to dry completely.
Cleaning the analyzer connector socket
Do not allow liquids to enter the connector socket. Wipe around, not over, the connector
socket.
INSTRUCTIONS
1. Use The connector socket can be cleaned by blowing compressed dry air or nitrogen (for
example, with a compressed invertible gas duster can) when required.
2. Guide the tube back and forth around the outside of the connector socket during
cleaning.
Do not insert the compressed air canister tube inside the connector socket.
Cleaning the analyzer cover and hinges
INSTRUCTIONS
1. Clean the cover and hinges with a cotton tip dampened with one of the following:
‒ 70% isopropanol alcohol
‒ 70% ethanol alcohol
‒ Antibacterial soap and water
2. Allow enough time after cleaning for the cover and hinges to dry completely.
Do not use cotton tips inside the cartridge slot.
Cleaning the analyzer cartridge insertion area
INSTRUCTIONS
1. Clean the analyzer insertion area with a cloth or wipe, dampened with one of the
following:
‒ 70% isopropanol alcohol