8. After analysis, the analyzer displays the result as a quantitative value along with Pass or
Fail.
9:11 AM 30-03-2020
QC Test Result
Patient 123456789
HScTnl
9:11 AM
Result
Pass
PTH
cTnI
39.1
5.1
pg / ml
Repeat
Finish
9. Do one of the following:
‒ Tap Finish.
‒ Tap the Result tile to view the detailed test results and then tap OK.
10. On the Finish screen, do one of the following:
‒ Tap Accept to accept the test results. The analyzer updates the cartridge lot locking
status depending on the result and configured workflow option.
‒ Tap Reject to not accept the test results.
11. Remove the cartridge and close the analyzer cover.
When QC guidance is turned on in the QC and Workflow settings dialog, failing to run
an overdue QC test locks the cartridge lot, preventing patient testing on that lot only.

Working with QC Test Results

Viewing QC test results

INSTRUCTIONS
> Results Review.
1. Tap
2. Do one of the following:
‒ To review specific QC test results and view test details: Tap a QC result on the Results
Review screen. Then tap the test result itself for additional details such as test date, Lot
ID, and expiry date.
‒ To navigate through the list of test results: Tap
3. Tap
to return to the Menu screen.
Filtering test results by date, test type, user, or cartridge type
INSTRUCTIONS
1. Tap
> Results Review > Filter.
2. Do one of the following:
‒ Tap Date to open the Filter Date screen.
• Choose to view results for Any Date, Today only, Last 7 Days, or you Specify Date
Range.
• For a specific date range, use the arrows on the Date Selection screen to select the
date range. Tap OK to confirm your selection.
‒ Tap Test to choose whether you want to view all tests, patient tests, or QC tests.
Chapter 3 - Performing Quality Control
.
28