Endress+Hauser Proline Cubemass C 300 Operating Instructions Manual

Endress+Hauser Proline Cubemass C 300 Operating Instructions Manual

Coriolis flowmeter

Advertisement

Quick Links

BA01483D/06/EN/05.22-00
71574975
2022-08-01
Valid as of version
01.06.zz (Device firmware)
Products
Operating Instructions
Proline Cubemass C 300
Coriolis flowmeter
HART
Solutions
Services

Advertisement

Table of Contents
loading

Summary of Contents for Endress+Hauser Proline Cubemass C 300

  • Page 1 Products Solutions Services BA01483D/06/EN/05.22-00 71574975 2022-08-01 Valid as of version 01.06.zz (Device firmware) Operating Instructions Proline Cubemass C 300 Coriolis flowmeter HART...
  • Page 2 • The manufacturer reserves the right to modify technical data without prior notice. Your Endress+Hauser Sales Center will supply you with current information and updates to these instructions. Endress+Hauser...
  • Page 3 Proline Cubemass C 300 Table of contents Table of contents About this document ....6 Mounting ......21 Document function .
  • Page 4 Table of contents Proline Cubemass C 300 8.3.12 Enabling and disabling the keypad 10.5.7 Configuration management ..124 lock ......
  • Page 5 13.2 Measuring and test equipment ... 171 13.3 Endress+Hauser services ....171 Repair ......172 14.1 General information .
  • Page 6 About this document Proline Cubemass C 300 About this document Document function These Operating Instructions contain all the information required in the various life cycle phases of the device: from product identification, incoming acceptance and storage, to installation, connection, operation and commissioning, through to troubleshooting, maintenance and disposal.
  • Page 7 Proline Cubemass C 300 About this document Symbol Meaning Light emitting diode is on. Light emitting diode is flashing. 1.2.4 Tool symbols Symbol Meaning Flat-blade screwdriver Allen key Open-ended wrench 1.2.5 Symbols for certain types of information Symbol Meaning Permitted Procedures, processes or actions that are permitted.
  • Page 8 For an overview of the scope of the associated Technical Documentation, refer to the following: • W@M Device Viewer (www.endress.com/deviceviewer): Enter the serial number from the nameplate • Endress+Hauser Operations App: Enter the serial number from the nameplate or scan the matrix code on the nameplate 1.3.1 Document function...
  • Page 9 Proline Cubemass C 300 Safety instructions Safety instructions Requirements for the personnel The personnel for installation, commissioning, diagnostics and maintenance must fulfill the following requirements: ‣ Trained, qualified specialists must have a relevant qualification for this specific function and task.
  • Page 10 Verification for borderline cases: ‣ For special fluids and fluids for cleaning, Endress+Hauser is glad to provide assistance in verifying the corrosion resistance of fluid-wetted materials, but does not accept any warranty or liability as minute changes in the temperature, concentration or level of contamination in the process can alter the corrosion resistance properties.
  • Page 11 (Statutory Instruments). These are listed in the UKCA Declaration of Conformity along with the designated standards. By selecting the order option for UKCA marking, Endress+Hauser confirms a successful evaluation and testing of the device by affixing the UKCA mark. Contact address Endress+Hauser UK: Endress+Hauser Ltd.
  • Page 12 Safety instructions Proline Cubemass C 300 (DIP switch on the main electronics module). When hardware write protection is enabled, only read access to the parameters is possible. Hardware write protection is disabled when the device is delivered →  131.
  • Page 13 Proline Cubemass C 300 Safety instructions 2.7.3 Access via Web server The device can be operated and configured via a Web browser with the integrated Web server (→  56). The connection is via the service interface (CDI-RJ45) or the WLAN interface.
  • Page 14 Product description Proline Cubemass C 300 Product description The device consists of a transmitter and a sensor. The device is available as a compact version: The transmitter and sensor form a mechanical unit. Product design N ic te r e rg ö...
  • Page 15 Is the envelope present with accompanying documents? • If one of the conditions is not satisfied, contact your Endress+Hauser Sales Center. • The Technical Documentation is available via the Internet or via the Endress+Hauser Operations App, see the "Product identification" section →  16.
  • Page 16 • The chapters "Additional standard documentation on the device" and "Supplementary device-dependent documentation" • The Device Viewer: Enter the serial number from the nameplate (www.endress.com/deviceviewer) • The Endress+Hauser Operations App: Enter the serial number from the nameplate or scan the DataMatrix code on the nameplate. 4.2.1 Transmitter nameplate...
  • Page 17 Proline Cubemass C 300 Incoming acceptance and product identification 4.2.2 Sensor nameplate 3 4 5 6 Order code: Ser. no.: Ext. ord. cd.: Size: Ptest = Material: 0044 Patents Date: 322541-0000 13 14 A0013907  3 Example of a sensor nameplate...
  • Page 18 Incoming acceptance and product identification Proline Cubemass C 300 4.2.3 Symbols on measuring device Symbol Meaning WARNING! This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in serious or fatal injury. To determine the nature of the potential hazard and the measures required to avoid it, consult the documentation accompanying the measuring device.
  • Page 19 Proline Cubemass C 300 Storage and transport Storage and transport Storage conditions Observe the following notes for storage: ‣ Store in the original packaging to ensure protection from shock. ‣ Do not remove protective covers or protective caps installed on process connections.
  • Page 20 Storage and transport Proline Cubemass C 300 5.2.2 Measuring devices with lifting lugs CAUTION Special transportation instructions for devices with lifting lugs ‣ Only use the lifting lugs fitted on the device or flanges to transport the device. ‣ The device must always be secured at two lifting lugs at least.
  • Page 21 Proline Cubemass C 300 Mounting Mounting Mounting requirements 6.1.1 Mounting position Mounting location A0028772 To prevent measuring errors arising from accumulation of gas bubbles in the measuring pipe, avoid the following mounting locations in the piping: • Highest point of a pipeline.
  • Page 22 Mounting Proline Cubemass C 300 Ø orifice plate, pipe restriction [mm] [in] [mm] [in] ¹⁄₂₄ 0.03 ¹⁄₁₂ 0.06 ¹⁄₈ 0.12 ¹⁄₄ 0.20 Orientation The direction of the arrow on the sensor nameplate helps you to install the sensor according to the flow direction (direction of medium flow through the piping).
  • Page 23 If operating outdoors: Avoid direct sunlight, particularly in warm climatic regions. You can order a weather protection cover from Endress+Hauser. →  174. System pressure It is important that cavitation does not occur, or that gases entrained in the liquids do not outgas.
  • Page 24 Mounting Proline Cubemass C 300 A0034391  5 Thermal insulation with not isolated extended neck Heating NOTICE Electronics can overheat due to elevated ambient temperature! ‣ Observe maximum permitted ambient temperature for the transmitter. ‣ Depending on the medium temperature, take the device orientation requirements into account.
  • Page 25 Proline Cubemass C 300 Mounting Rupture disk Process-related information: →  195. WARNING Danger from medium escaping! Medium escaping under pressure can cause injury or material damage. ‣ Take precautions to prevent danger to persons and damage if the rupture disk is actuated.
  • Page 26 Mounting Proline Cubemass C 300 Horizontal Device standing on a solid support base A0030287 Mounting plate The universal mounting plate can be used to affix or place the unit on a flat surface (order code for "Accessories", option PA). 8 mm A0019768 ...
  • Page 27 Proline Cubemass C 300 Mounting Mounting the measuring device 6.2.1 Required tools For sensor For flanges and other process connections: use a suitable mounting tool 6.2.2 Preparing the measuring device 1. Remove all remaining transport packaging. 2. Remove any protective covers or protective caps present from the sensor.
  • Page 28 Mounting Proline Cubemass C 300 – – 3 mm 4 mm 7 Nm (5.2 lbf ft) A0029993  8 Non-Ex housing 1. Depending on the device version: Loosen the securing clamp of the connection compartment cover. 2. Unscrew the connection compartment cover.
  • Page 29 Proline Cubemass C 300 Mounting – – – 3 mm A0030035 1. Depending on the device version: Loosen the securing clamp of the connection compartment cover. 2. Unscrew the connection compartment cover. 3. Turn the display module to the desired position: max. 8 × 45° in each direction.
  • Page 30 Electrical connection Proline Cubemass C 300 Electrical connection WARNING Live parts! Incorrect work performed on the electrical connections can result in an electric shock. ‣ Set up a disconnecting device (switch or power-circuit breaker) to easily disconnect the device from the supply voltage.
  • Page 31 Proline Cubemass C 300 Electrical connection Double pulse output Standard installation cable is sufficient Relay output Standard installation cable is sufficient. Current input 0/4 to 20 mA Standard installation cable is sufficient Status input Standard installation cable is sufficient Cable diameter •...
  • Page 32 Electrical connection Proline Cubemass C 300 Cable length Maximum 300 m (1 000 ft), maximum loop impedance 20 Ω Capacitance: core/shield Maximum 1 000 nF for Zone 1, Class I, Division 1 Maximum 24 µH/Ω for Zone 1, Class I, Division 1...
  • Page 33 Proline Cubemass C 300 Electrical connection 7.2.3 Terminal assignment Transmitter: supply voltage, input/outputs The terminal assignment of the inputs and outputs depends on the individual order version of the device. The device-specific terminal assignment is documented on an adhesive label in the terminal cover.
  • Page 34 Electrical connection Proline Cubemass C 300 – – – 3 mm A0029813 1. Loosen the securing clamp of the connection compartment cover. 2. Unscrew the connection compartment cover. 3. Squeeze the tabs of the display module holder together. 4. Remove the display module holder.
  • Page 35 Proline Cubemass C 300 Electrical connection 22 mm 24 mm – A0029816 10. Connect the cable according to the terminal assignment.  Signal cable terminal assignment: The device-specific terminal assignment is documented on an adhesive label in the terminal cover.
  • Page 36 Electrical connection Proline Cubemass C 300 7.3.2 Connecting the remote display and operating module DKX001 The remote display and operating module DKX001 is available as an optional extra →  174. • The measuring device is always supplied with a dummy cover when the remote display and operating module DKX001 is ordered directly with the measuring device.
  • Page 37 Proline Cubemass C 300 Electrical connection Special connection instructions 7.5.1 Connection examples Current output 4 to 20 mA HART 4...20 mA A0029055  11 Connection example for 4 to 20 mA HART current output (active) Automation system with current input (e.g. PLC) Cable shield provided at one end.
  • Page 38 Electrical connection Proline Cubemass C 300 HART input 4...20 mA A0028763  13 Connection example for HART input with a common negative (passive) Automation system with HART output (e.g. PLC) Active barrier for power supply (e.g. RN221N) Cable shield provided at one end. The cable shield must be grounded at both ends to comply with EMC requirements;...
  • Page 39 Proline Cubemass C 300 Electrical connection Pulse/frequency output 12345 A0028761  16 Connection example for pulse/frequency output (passive) Automation system with pulse/frequency input (e.g. PLC with 10 kΩ pull-up or pull-down resistor) Power supply Transmitter: observe input values →  183...
  • Page 40 Electrical connection Proline Cubemass C 300 A0029279  19 Connection example for double pulse output (passive) Automation system with double pulse input (e.g. PLC with a 10 kΩ pull-up or pull-down resistor) Power supply Transmitter: observe input values →  184...
  • Page 41 Proline Cubemass C 300 Electrical connection Status input A0028764  22 Connection example for status input Automation system with status output (e.g. PLC) Power supply Transmitter Ensuring the degree of protection The measuring device fulfills all the requirements for the degree of protection IP66/67, Type 4X enclosure.
  • Page 42 Electrical connection Proline Cubemass C 300 If supply voltage is present, do values appear on the display module?  Are dummy plugs inserted in unused cable entries and have transportation plugs been replaced with dummy plugs? Endress+Hauser...
  • Page 43 Proline Cubemass C 300 Operation options Operation options Overview of operation options A0034513 Local operation via display module Computer with Web browser (e.g. Internet Explorer) or with operating tool (e.g. FieldCare, DeviceCare, AMS Device Manager, SIMATIC PDM) Field Xpert SFX350 or SFX370...
  • Page 44 Operation options Proline Cubemass C 300 Structure and function of the operating menu 8.2.1 Structure of the operating menu For an overview of the operating menu for experts: see the "Description of Device Parameters" document supplied with the device Operating menu for operators and maintenances...
  • Page 45 Proline Cubemass C 300 Operation options 8.2.2 Operating philosophy The individual parts of the operating menu are assigned to certain user roles (operator, maintenance etc.). Each user role contains typical tasks within the device lifecycle. Menu/parameter User role and tasks...
  • Page 46 Operation options Proline Cubemass C 300 Access to the operating menu via the local display 8.3.1 Operational display X X X X X X X 1120.50 kg/h A0029348 Operational display Device tag Status area Display area for measured values (4-line) Operating elements →...
  • Page 47 Proline Cubemass C 300 Operation options • Density • Reference density Temperature Totalizer  The measurement channel number indicates which of the three totalizers is displayed. Output  The measurement channel number indicates which of the outputs is displayed. Status input...
  • Page 48 Operation options Proline Cubemass C 300 • In the submenu: Omission symbol for Name of current Display symbol for menu operating menu levels in • Submenu • In the wizard: between • Wizard Display symbol for wizard • Parameters ↓...
  • Page 49 Proline Cubemass C 300 Operation options Locking Symbol Meaning Parameter locked When displayed in front of a parameter name, indicates that the parameter is locked. • By a user-specific access code • By the hardware write protection switch Wizard operation...
  • Page 50 Operation options Proline Cubemass C 300 Text editor Xxxxxxx A B C D E F G H I J K M N O P Q R S T U V W A0034114  25 For entering text in parameters (e.g. tag name)
  • Page 51 Proline Cubemass C 300 Operation options Controlling data entries Symbol Meaning Move entry position Reject entry Confirm entry Delete character immediately to the left of the entry position Delete character immediately to the right of the entry position Clear all the characters entered 8.3.4...
  • Page 52 Operation options Proline Cubemass C 300 Meaning Escape key combination (press keys simultaneously) In menu, submenu • Pressing the key briefly: • Exits the current menu level and takes you to the next level up. • If help text is open, closes the help text of the parameter.
  • Page 53 Proline Cubemass C 300 Operation options 8.3.6 Navigating and selecting from list Different operating elements are used to navigate through the operating menu. The navigation path is displayed on the left in the header. Icons are displayed in front of the individual menus.
  • Page 54 Operation options Proline Cubemass C 300 The direct access code consists of a 5-digit number (at maximum) and the channel number, which identifies the channel of a process variable: e.g. 00914-2. In the navigation view, this appears on the right-hand side in the header of the selected parameter.
  • Page 55 Proline Cubemass C 300 Operation options Ent. access code Invalid or out of range input value Min:0 Max:9999 A0014049-EN For a description of the editing view - consisting of the text editor and numeric editor - with symbols →  49, for a description of the operating elements →  51 8.3.10...
  • Page 56 Operation options Proline Cubemass C 300 2. Enter the access code.  The -symbol in front of the parameters disappears; all previously write- protected parameters are now re-enabled. 8.3.12 Enabling and disabling the keypad lock The keypad lock makes it possible to block access to the entire operating menu via local operation.
  • Page 57 Proline Cubemass C 300 Operation options 8.4.2 Requirements Computer hardware Hardware Interface CDI-RJ45 WLAN Interface The computer must have an RJ45 The operating unit must have a interface. WLAN interface. Connection Ethernet cable with RJ45 connector. Connection via Wireless LAN.
  • Page 58 Operation options Proline Cubemass C 300 Measuring device: Via CDI-RJ45 service interface Device CDI-RJ45 service interface Measuring device The measuring device has an RJ45 interface. Web server Web server must be enabled; factory setting: ON  For information on enabling the Web server →  62...
  • Page 59 Proline Cubemass C 300 Operation options Via WLAN interface Configuring the Internet protocol of the mobile terminal NOTICE If the WLAN connection is lost during the configuration, settings made may be lost. ‣ Make sure that the WLAN connection is not disconnected while configuring the device.
  • Page 60 Operation options Proline Cubemass C 300 2. Enter the IP address of the web server in the address line of the web browser: 192.168.1.212  The login page appears. 2 3 4 A0029417 Picture of device Device name Device tag (→  79)
  • Page 61 Proline Cubemass C 300 Operation options 8.4.5 User interface A0029418 Function row Local display language Navigation area Header The following information appears in the header: • Device name • Device tag • Device status with status signal →  155 •...
  • Page 62 Operation options Proline Cubemass C 300 Working area Depending on the selected function and the related submenus, various actions can be performed in this area: • Configuring parameters • Reading measured values • Calling up help text • Starting an upload/download 8.4.6...
  • Page 63 Proline Cubemass C 300 Operation options Access to the operating menu via the operating tool The structure of the operating menu in the operating tools is the same as for operation via the local display. 8.5.1 Connecting the operating tool Via HART protocol This communication interface is available in device versions with a HART output.
  • Page 64 Operation options Proline Cubemass C 300 A0028746  28 Options for remote operation via HART protocol (passive) Control system (e.g. PLC) Transmitter power supply unit, e.g. RN221N (with communication resistor) Connection for Commubox FXA195 and Field Communicator 475 Field Communicator 475 Computer with Web browser (e.g.
  • Page 65 Proline Cubemass C 300 Operation options Via WLAN interface The optional WLAN interface is available on the following device version: Order code for "Display; operation", option G "4-line, illuminated; touch control + WLAN" A0034570 Transmitter with integrated WLAN antenna Transmitter with external WLAN antenna...
  • Page 66 FieldCare Function scope FDT-based plant asset management tool from Endress+Hauser. It can configure all smart field devices in a system and helps you manage them. By using the status information, it is also a simple but effective way of checking their status and condition.
  • Page 67 Proline Cubemass C 300 Operation options Access is via: • HART protocol • CDI-RJ45 service interface →  64 • WLAN interface →  65 Typical functions: • Parameterization of transmitters • Loading and saving device data (upload/download) • Documentation of the measuring point •...
  • Page 68 DeviceCare Function scope Tool for connecting and configuring Endress+Hauser field devices. The fastest way to configure Endress+Hauser field devices is with the dedicated "DeviceCare" tool. Together with the device type managers (DTMs) it presents a convenient, comprehensive solution. For details, see Innovation Brochure IN01047S Source for device description files See information →...
  • Page 69 Proline Cubemass C 300 Operation options Source for device description files See data →  70 8.5.6 SIMATIC PDM Function scope SIMATIC PDM is a standardized, manufacturer-independent program from Siemens for the operation, configuration, maintenance and diagnosis of intelligent field devices via HART protocol.
  • Page 70 System integration Proline Cubemass C 300 System integration Overview of device description files 9.1.1 Current version data for the device Firmware version 01.06.zz • On the title page of the Operating Instructions • On the transmitter nameplate • Firmware version Diagnostics →...
  • Page 71 Proline Cubemass C 300 System integration Measured variables via HART protocol The following measured variables (HART device variables) are assigned to the dynamic variables at the factory: Dynamic variables Measured variables (HART device variables) Primary dynamic variable (PV) Mass flow...
  • Page 72 System integration Proline Cubemass C 300 Measured variables for PV (primary dynamic variable) • Measured variables which are generally available: • Mass flow • Volume flow • Corrected volume flow • Density • Reference density • Temperature • Electronics temperature •...
  • Page 73 Proline Cubemass C 300 System integration Measured variables for SV, TV, QV (secondary, tertiary and quaternary dynamic variable) • Measured variables which are always available: • Mass flow • Volume flow • Corrected volume flow • Density • Reference density •...
  • Page 74 System integration Proline Cubemass C 300 Other settings Burst mode functionality in accordance with HART 7 Specification: Navigation "Expert" menu → Communication → HART output → Burst configuration → Burst configuration 1 to n ‣ Burst configuration 1 to n Burst mode 1 to n →...
  • Page 75 Proline Cubemass C 300 System integration Parameter Description Selection / User entry Factory setting Burst variable 0 For HART command 9 and 33: select the • Mass flow Volume flow HART device variable or the process variable. • Volume flow • Corrected volume flow •...
  • Page 76 System integration Proline Cubemass C 300 Parameter Description Selection / User entry Factory setting Min. update period Enter the minimum time span between two Positive integer 1 000 ms burst commands of burst message X. Max. update period Enter the maximum time span between two...
  • Page 77 Proline Cubemass C 300 Commissioning Commissioning 10.1 Function check Before commissioning the measuring device: ‣ Make sure that the post-installation and post-connection checks have been performed. • Checklist for "Post-mounting check" →  29 • Checklist for "Post-connection check" →  41 10.2...
  • Page 78 Commissioning Proline Cubemass C 300 X X X X X X X 20.50 0104-1 Main menu Display language English Display/operat. Setup Main menu Display/operat. Setup Diagnostic / ../Setup Medium selection XXXXXXXXX XXXXXXXXX A0032222-EN  31 Taking the example of the local display The number of submenus and parameters can vary depending on the device version.
  • Page 79 Proline Cubemass C 300 Commissioning ‣ Partially filled pipe detection →  107 ‣ Advanced setup →  108 10.4.1 Defining the tag name To enable fast identification of the measuring point within the system, you can enter a unique designation using the Device tag parameter and thus change the factory setting.
  • Page 80 Commissioning Proline Cubemass C 300 Corrected volume flow unit →  80 Corrected volume unit →  80 →  80 Density unit Reference density unit →  80 Density 2 unit →  81 Temperature unit →  81 →...
  • Page 81 Proline Cubemass C 300 Commissioning Parameter Description Selection Factory setting Density 2 unit Select second density unit. Unit choose list Depends on country: • kg/l • lb/ft³ Temperature unit Select temperature unit. Unit choose list Country-specific: • °C Effect • °F The selected unit applies for: •...
  • Page 82 Commissioning Proline Cubemass C 300 10.4.3 Selecting and setting the medium The Select medium wizard submenu contains parameters that must be configured in order to select and set the medium. Navigation "Setup" menu → Medium selection ‣ Medium selection MFT (Multi-Frequency Technology) Select medium type →...
  • Page 83 Proline Cubemass C 300 Commissioning Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry / User interface Select medium type – Use this function to select the • Liquid Liquid type of medium: "Gas" or • Gas "Liquid". Select the "Other"...
  • Page 84 Commissioning Proline Cubemass C 300 Parameter Prerequisite Description Selection / User Factory setting entry / User interface Pressure value In the Pressure compensation Enter process pressure to be Positive floating- 1.01325 bar parameter, the Fixed value used for pressure correction. point number option is selected.
  • Page 85 Proline Cubemass C 300 Commissioning Parameter Description User interface / Selection /  Factory setting User entry Apply I/O configuration Apply parameterization of the freely • No configurable I/O module. • Yes I/O alteration code Enter the code in order to change the I/O Positive integer configuration.
  • Page 86 Commissioning Proline Cubemass C 300 Parameter Prerequisite Description User interface /  Factory setting Selection / User entry Current span – Select current range for • 4...20 mA (4... Country-specific: process value output and 20.5 mA) • 4...20 mA NE upper/lower level for alarm •...
  • Page 87 Proline Cubemass C 300 Commissioning Parameter overview with brief description Parameter Description Selection / User interface /  Factory setting User entry Assign status input Select function for the status input. • Off • Reset totalizer 1 • Reset totalizer 2 • Reset totalizer 3 •...
  • Page 88 Commissioning Proline Cubemass C 300 Parameter overview with brief description Parameter Prerequisite Description User interface /  Factory setting Selection / User entry Terminal number – Shows the terminal numbers • Not used – used by the current output • 26-27 (I/O 1) module.
  • Page 89 Proline Cubemass C 300 Commissioning Parameter Prerequisite Description User interface /  Factory setting Selection / User entry Process variable current output – Select the process variable for • Off Mass flow the current output. • Mass flow • Volume flow • Corrected volume flow •...
  • Page 90 Commissioning Proline Cubemass C 300 Parameter Prerequisite Description User interface /  Factory setting Selection / User entry Current range output – Select current range for • 4...20 mA NE Depends on country: process value output and (3.8...20.5 mA) • 4...20 mA NE upper/lower level for alarm •...
  • Page 91 Proline Cubemass C 300 Commissioning 10.4.8 Configuring the pulse/frequency/switch output The Pulse/frequency/switch output wizard guides you systematically through all the parameters that can be set for configuring the selected output type. Navigation "Setup" menu → Advanced setup → Pulse/frequency/switch output ‣...
  • Page 92 Commissioning Proline Cubemass C 300 Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting interface / User entry Operating mode – Define the output as a pulse, • Pulse Pulse frequency or switch output. • Frequency • Switch Terminal number –...
  • Page 93 Proline Cubemass C 300 Commissioning Configuring the frequency output Navigation "Setup" menu → Pulse/frequency/switch output ‣ Pulse/frequency/switch output 1 to n Operating mode →  93 Terminal number →  93 Signal mode →  93 Assign frequency output →  94 Minimum frequency value →...
  • Page 94 Commissioning Proline Cubemass C 300 Parameter Prerequisite Description Selection / User Factory setting interface / User entry Assign frequency output In the Operating mode Select process variable for • Off parameter (→  91), the frequency output. • Mass flow Frequency option is selected.
  • Page 95 Proline Cubemass C 300 Commissioning Parameter Prerequisite Description Selection / User Factory setting interface / User entry Maximum frequency value The Frequency option is Enter maximum frequency. 0.0 to 10 000.0 Hz 10 000.0 Hz selected in the Operating mode parameter (→  91)
  • Page 96 Commissioning Proline Cubemass C 300 Configuring the switch output Navigation "Setup" menu → Pulse/frequency/switch output ‣ Pulse/frequency/switch output 1 to n Operating mode →  96 Terminal number →  96 Signal mode →  96 Switch output function →  97 Assign diagnostic behavior →...
  • Page 97 Proline Cubemass C 300 Commissioning Parameter Prerequisite Description Selection / User Factory setting interface / User entry Switch output function The Switch option is selected Select function for switch • Off in the Operating mode output. • On parameter. • Diagnostic behavior • Limit •...
  • Page 98 Commissioning Proline Cubemass C 300 Parameter Prerequisite Description Selection / User Factory setting interface / User entry Switch-on value • The Switch option is Enter measured value for the Signed floating-point Country-specific: selected in the Operating switch-on point. number • 0 kg/h mode parameter.
  • Page 99 Proline Cubemass C 300 Commissioning Switch-off value →  100 Switch-off delay →  100 →  100 Switch-on value Switch-on delay →  100 Failure mode →  100 Parameter overview with brief description Parameter Prerequisite Description User interface / ...
  • Page 100 Commissioning Proline Cubemass C 300 Parameter Prerequisite Description User interface /  Factory setting Selection / User entry Assign limit The Limit option is selected in Select process variable for limit • Mass flow Mass flow the Relay output function function. • Volume flow parameter.
  • Page 101 Proline Cubemass C 300 Commissioning Navigation "Setup" menu → Double pulse output ‣ Double pulse output Signal mode →  101 Master terminal number →  101 →  101 Assign pulse output Measuring mode →  101 Value per pulse →...
  • Page 102 Commissioning Proline Cubemass C 300 Parameter Description Selection / User interface /  Factory setting User entry Failure mode Define output behavior in alarm condition. • Actual value No pulses • No pulses Invert output signal Invert the output signal. • No • Yes Visibility depends on order options or device settings 10.4.11 Configuring the local display...
  • Page 103 Proline Cubemass C 300 Commissioning Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry Format display A local display is provided. Select how measured values • 1 value, max. size 1 value, max. size are shown on the display.
  • Page 104 Commissioning Proline Cubemass C 300 Parameter Prerequisite Description Selection / User Factory setting entry Value 1 display A local display is provided. Select the measured value that • Mass flow Mass flow is shown on the local display. • Volume flow • Corrected volume flow •...
  • Page 105 Proline Cubemass C 300 Commissioning Parameter Prerequisite Description Selection / User Factory setting entry 100% bargraph value 1 A local display is provided. Enter 100% value for bar Signed floating-point Depends on country graph display. number and nominal diameter Value 2 display A local display is provided.
  • Page 106 Commissioning Proline Cubemass C 300 10.4.12 Configuring the low flow cut off The Low flow cut off wizard systematically guides the user through all the parameters that must be set to configure low flow cut off. Navigation "Setup" menu → Low flow cut off ‣...
  • Page 107 Proline Cubemass C 300 Commissioning 10.4.13 Configuring partially filled pipe detection The Partial filled pipe detection wizard guides you systematically through all parameters that have to be set for configuring the monitoring of the pipe filling. Navigation "Setup" menu → Partially filled pipe detection ‣...
  • Page 108 Commissioning Proline Cubemass C 300 10.5 Advanced settings The Advanced setup submenu with its submenus contains parameters for specific settings. Navigation to the "Advanced setup" submenu X X X X X X X 20.50 0104-1 Main menu Display language English Display/operat.
  • Page 109: Table Of Contents

    Proline Cubemass C 300 Commissioning ‣ Sensor adjustment →  110 ‣ Totalizer 1 to n →  114 ‣ →  116 Display ‣ WLAN settings →  122 ‣ Configuration backup →  124 ‣ Administration →  125 10.5.1...
  • Page 110: Sensor Adjustment

    Commissioning Proline Cubemass C 300 Reference temperature (1816) →  110 Linear expansion coefficient (1817) →  110 →  110 Square expansion coefficient (1818) Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting interface / User entry Select reference density –...
  • Page 111 Proline Cubemass C 300 Commissioning Navigation "Setup" menu → Advanced setup → Sensor adjustment ‣ Sensor adjustment Installation direction →  111 ‣ Zero verification →  111 ‣ →  113 Zero adjustment Parameter overview with brief description Parameter...
  • Page 112 Commissioning Proline Cubemass C 300 Navigation "Setup" menu → Advanced setup → Sensor adjustment → Zero verification ‣ Zero verification Process conditions →  112 Progress →  112 →  112 Status Additional information →  112 Recommendation: →  112 Root cause →...
  • Page 113 Proline Cubemass C 300 Commissioning Parameter Description Selection / User interface Factory setting Zero point measured Shows the zero point measured for the Signed floating-point number – adjustment. Zero point standard deviation Shows the standard deviation of the zero Positive floating-point number –...
  • Page 114: Totalizer 1 To N

    Commissioning Proline Cubemass C 300 Parameter overview with brief description Parameter Description Selection / User interface Factory setting Process conditions Ensure process conditions as follows. • Tubes are completely filled – • Process operational pressure applied • No-flow conditions (closed valves) • Process and ambient...
  • Page 115 Proline Cubemass C 300 Commissioning Parameter overview with brief description Parameter Prerequisite Description Selection Factory setting Assign process variable – Select process variable for • Off Mass flow totalizer. • Mass flow • Volume flow • Corrected volume flow • Target mass flow •...
  • Page 116: Display

    Commissioning Proline Cubemass C 300 10.5.5 Carrying out additional display configurations In the Display submenu you can set all the parameters associated with the configuration of the local display. Navigation "Setup" menu → Advanced setup → Display ‣ Display Format display →...
  • Page 117 Proline Cubemass C 300 Commissioning 0% bargraph value 7 →  121 →  121 100% bargraph value 7 Decimal places 7 →  121 Value 8 display →  121 Decimal places 8 →  121 Display language →  122 Display interval →...
  • Page 118 Commissioning Proline Cubemass C 300 Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry Format display A local display is provided. Select how measured values • 1 value, max. size 1 value, max. size are shown on the display.
  • Page 119 Proline Cubemass C 300 Commissioning Parameter Prerequisite Description Selection / User Factory setting entry Value 1 display A local display is provided. Select the measured value that • Mass flow Mass flow is shown on the local display. • Volume flow • Corrected volume flow •...
  • Page 120 Commissioning Proline Cubemass C 300 Parameter Prerequisite Description Selection / User Factory setting entry 100% bargraph value 1 A local display is provided. Enter 100% value for bar Signed floating-point Depends on country graph display. number and nominal diameter Decimal places 1...
  • Page 121 Proline Cubemass C 300 Commissioning Parameter Prerequisite Description Selection / User Factory setting entry Decimal places 5 A measured value is specified Select the number of decimal • x x.xx in the Value 5 display places for the display value. • x.x parameter.
  • Page 122: Wlan Settings

    Commissioning Proline Cubemass C 300 Parameter Prerequisite Description Selection / User Factory setting entry Display language A local display is provided. Set display language. • English English • Deutsch (alternatively, the • Français ordered language is • Español preset in the device) •...
  • Page 123 Proline Cubemass C 300 Commissioning Navigation "Setup" menu → Advanced setup → WLAN settings ‣ WLAN settings WLAN IP address →  123 Security type →  123 →  123 WLAN passphrase Assign SSID name →  123 SSID name →...
  • Page 124: Configuration Backup

    Commissioning Proline Cubemass C 300 10.5.7 Configuration management After commissioning, you can save the current device configurationor restore the previous device configuration. You can do so using the Configuration management parameter and the related options found in the Configuration backup submenu.
  • Page 125: Administration

    Proline Cubemass C 300 Commissioning Function scope of the "Configuration management" parameter Options Description Cancel No action is executed and the user exits the parameter. Execute backup A backup copy of the current device configuration is saved from the HistoROM backup to the memory of the device.
  • Page 126: Parameter Overview With Brief Description

    Commissioning Proline Cubemass C 300 Parameter overview with brief description Parameter Description User entry Define access code Restrict write-access to parameters to protect the configuration Max. 16-digit character string comprising of the device against unintentional changes. numbers, letters and special characters Confirm access code Confirm the entered access code.
  • Page 127 Proline Cubemass C 300 Commissioning 10.6 Simulation The Simulation submenu enables you to simulate, without a real flow situation, various process variables in the process and the device alarm mode and to verify downstream signal chains (switching valves or closed-control loops).
  • Page 128 Commissioning Proline Cubemass C 300 Status input 1 to n simulation →  129 Input signal level 1 to n →  129 Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry Assign simulation process variable –...
  • Page 129 Proline Cubemass C 300 Commissioning Parameter Prerequisite Description Selection / User Factory setting entry Relay output 1 to n simulation – Switch simulation of the relay • Off output on and off. • On Switch state 1 to n The On option is selected in Select status of the relay •...
  • Page 130 Commissioning Proline Cubemass C 300 Defining the access code via local display 1. Navigate to the Define access code parameter (→  126). 2. Define a max. 16-digit character string comprising numbers, letters and special characters as the access code.
  • Page 131 Commissioning Via Web browser, FieldCare, DeviceCare (via CDI-RJ45 service interface), fieldbus You can only obtain a reset code from your local Endress+Hauser service organization. The code must be calculated explicitly for every device. 1. Note down the serial number of the device.
  • Page 132 Commissioning Proline Cubemass C 300 2. Setting the write protection (WP) switch on the main electronics module to the OFF position (factory setting) disables hardware write protection.  No option is displayed in the Locking status parameter →  133. On the local display, the ...
  • Page 133 Proline Cubemass C 300 Operation Operation 11.1 Reading off the device locking status Device active write protection: Locking status parameter Operation → Locking status Function scope of the "Locking status" parameter Options Description None The access authorization displayed in the Access status parameter applies →...
  • Page 134 Operation Proline Cubemass C 300 11.4.1 "Measured variables" submenu The Measured variables submenu contains all the parameters needed to display the current measured values for each process variable. Navigation "Diagnostics" menu → Measured values → Measured variables ‣ Measured variables Mass flow →...
  • Page 135 Proline Cubemass C 300 Operation Parameter Prerequisite Description User interface Corrected volume flow – Displays the corrected volume flow that Signed floating-point is currently calculated. number Dependency The unit is taken from: Corrected volume flow unit parameter (→  80) Density –...
  • Page 136 Operation Proline Cubemass C 300 Parameter Prerequisite Description User interface Carrier corrected volume flow With the following conditions: Displays the corrected volume flow Signed floating-point • Order code for "Application package", currently measured for the carrier fluid. number option ED "Concentration"...
  • Page 137 Proline Cubemass C 300 Operation Parameter overview with brief description Parameter Prerequisite Description User interface Totalizer value 1 to n A process variable is selected in the Displays the current totalizer counter Signed floating-point Assign process variable parameter reading. number (→...
  • Page 138 Operation Proline Cubemass C 300 Navigation "Diagnostics" menu → Measured values → Input values → Status input 1 to n ‣ Status input 1 to n Value status input →  138 Parameter overview with brief description Parameter Description User interface Value status input Shows the current input signal level.
  • Page 139 Proline Cubemass C 300 Operation Parameter overview with brief description Parameter Description User interface Output current Displays the current value currently calculated for the current 3.59 to 22.5 mA output. Measured current Displays the current value currently measured for the current 0 to 30 mA output.
  • Page 140 Operation Proline Cubemass C 300 Parameter overview with brief description Parameter Description User interface Switch state Shows the current relay switch status. • Open • Closed Switch cycles Shows number of all performed switch cycles. Positive integer Max. switch cycles number Shows the maximal number of guaranteed switch cycles.
  • Page 141 Proline Cubemass C 300 Operation Totalizer value 1 to n →  141 Reset all totalizers →  141 Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry / User interface Control Totalizer 1 to n A process variable is selected Control totalizer value.
  • Page 142 Operation Proline Cubemass C 300 11.7 Show data logging The Extended HistoROM application package must be enabled in the device (order option) for the Data logging submenu to appear. This contains all the parameters for the measured value history. Data logging is also available via: •...
  • Page 143 Proline Cubemass C 300 Operation Data logging status →  145 →  145 Entire logging duration ‣ Display channel 1 ‣ Display channel 2 ‣ Display channel 3 ‣ Display channel 4 Endress+Hauser...
  • Page 144 Operation Proline Cubemass C 300 Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry / User interface Assign channel 1 The Extended HistoROM Assign process variable to • Off application package is logging channel. • Mass flow available. • Volume flow •...
  • Page 145 Proline Cubemass C 300 Operation Parameter Prerequisite Description Selection / User Factory setting entry / User interface Assign channel 2 The Extended HistoROM Assign a process variable to For the picklist, see application package is logging channel. the Assign channel available. 1 parameter (→  144) ...
  • Page 146 Operation Proline Cubemass C 300 In the case of two-phase media, the Gas Fraction Handler stabilizes the output values and enables better readability for operators and easier interpretation by the distributed control system. The level of smoothing is adjusted according to the severity of the disturbances introduced by the second phase.
  • Page 147 Proline Cubemass C 300 Operation Parameter overview with brief description Parameter Prerequisite Description User interface / User Factory setting entry Inhomogeneous medium index – Shows the degree of Signed floating-point – inhomogeneity of the medium. number Cut off inhomogeneous wet gas –...
  • Page 148 Diagnostics and troubleshooting Proline Cubemass C 300 Diagnostics and troubleshooting 12.1 General troubleshooting For local display Error Possible causes Remedy Local display dark and no output Supply voltage does not match that Apply the correct supply voltage signals specified on the nameplate.
  • Page 149 Proline Cubemass C 300 Diagnostics and troubleshooting Error Possible causes Remedial action Device shows correct value on local Parametrization errors Check parameterization and correct display, but signal output is incorrect, though in the valid range. Device measures incorrectly. Configuration error or device is 1.
  • Page 150 Diagnostics and troubleshooting Proline Cubemass C 300 Error Possible causes Remedy Connection lost 1. Check cable connection and power supply. 2. Refresh the Web browser and restart if necessary. Content of Web browser incomplete or difficult Not using optimum version of Web server.
  • Page 151 Proline Cubemass C 300 Diagnostics and troubleshooting Color Meaning Flashing red/green The device restarts. Device status (during Flashes red slowly If > 30 seconds: problem with the boot loader. start-up) Flashes red quickly If > 30 seconds: compatibility problem when reading the firmware.
  • Page 152 Diagnostics and troubleshooting Proline Cubemass C 300 12.3 Diagnostic information on local display 12.3.1 Diagnostic message Faults detected by the self-monitoring system of the measuring device are displayed as a diagnostic message in alternation with the operational display. Operational display in alarm condition...
  • Page 153 Proline Cubemass C 300 Diagnostics and troubleshooting Diagnostic behavior Symbol Meaning Alarm • Measurement is interrupted. • Signal outputs and totalizers assume the defined alarm condition. • A diagnostic message is generated. Warning Measurement is resumed. The signal outputs and totalizers are not affected. A diagnostic message is generated.
  • Page 154 Diagnostics and troubleshooting Proline Cubemass C 300 12.3.2 Calling up remedial measures X X X X X X X X X X X X X X 20.50 S801 Supply voltage Menu Diagnostic list Diagnostics 1 S801 Supply voltage Diagnostics 2...
  • Page 155 Proline Cubemass C 300 Diagnostics and troubleshooting A0031056 Status area with status signal Diagnostic information Remedial measures with service ID In addition, diagnostic events which have occurred can be shown in the Diagnostics menu: • Via parameter →  163 •...
  • Page 156 Diagnostics and troubleshooting Proline Cubemass C 300 12.5 Diagnostic information in FieldCare or DeviceCare 12.5.1 Diagnostic options Any faults detected by the measuring device are displayed on the home page of the operating tool once the connection has been established.
  • Page 157 Proline Cubemass C 300 Diagnostics and troubleshooting 12.5.2 Calling up remedy information Remedy information is provided for every diagnostic event to ensure that problems can be rectified quickly: • On the home page Remedy information is displayed in a separate field below the diagnostics information.
  • Page 158 Diagnostics and troubleshooting Proline Cubemass C 300 Available status signals Configuration as per HART 7 Specification (Condensed Status), in accordance with NAMUR NE107. Symbol Meaning Failure A device error is present. The measured value is no longer valid. A0013956 Function check The device is in service mode (e.g.
  • Page 159 Proline Cubemass C 300 Diagnostics and troubleshooting Diagnostic Short text Remedy instructions Status Diagnostic number signal behavior [from the [from the factory] factory] Sensor initialization Sensor initialization in progress, please Warning active wait Sensor signal 1. If available: Check connection cable...
  • Page 160 Diagnostics and troubleshooting Proline Cubemass C 300 Diagnostic Short text Remedy instructions Status Diagnostic number signal behavior [from the [from the factory] factory] Firmware update failed 1. Update firmware of device Warning 2. Restart device Writing in HistoROM 1. Replace user interface board...
  • Page 161 Proline Cubemass C 300 Diagnostics and troubleshooting Diagnostic Short text Remedy instructions Status Diagnostic number signal behavior [from the [from the factory] factory] Current input 1 to n 1. Check process Warning faulty 2. Check current input settings Flow override active...
  • Page 162 Diagnostics and troubleshooting Proline Cubemass C 300 Diagnostic Short text Remedy instructions Status Diagnostic number signal behavior [from the [from the factory] factory] Custody transfer 1. Deactivate custody transfer mode Warning logbook full 2. Clear custody transfer logbook (all 30 entries) 3.
  • Page 163 Proline Cubemass C 300 Diagnostics and troubleshooting Diagnostic Short text Remedy instructions Status Diagnostic number signal behavior [from the [from the factory] factory] Oscillation damping too Check process conditions Warning high Condensation risk 1. Decrease ambient temperature Warning 2. Increase medium temperature Diagnostic behavior can be changed.
  • Page 164 Diagnostics and troubleshooting Proline Cubemass C 300 Parameter Prerequisite Description User interface Operating time from restart – Shows the time the device has been in Days (d), hours (h), operation since the last device restart. minutes (m) and seconds Operating time –...
  • Page 165 Proline Cubemass C 300 Diagnostics and troubleshooting In addition to the operation time of its occurrence, each event is also assigned a symbol that indicates whether the event has occurred or is ended: • Diagnostic event • : Occurrence of the event •...
  • Page 166 Diagnostics and troubleshooting Proline Cubemass C 300 Info number Info name I1222 Zero point adjustment ok I1256 Display: access status changed I1264 Safety sequence aborted I1278 I/O module restarted I1335 Firmware changed I1361 Web server: login failed I1397 Fieldbus: access status changed...
  • Page 167 Proline Cubemass C 300 Diagnostics and troubleshooting Info number Info name I1635 Reset to delivery settings I1639 Max. switch cycles number reached I1643 Custody transfer logbook cleared I1649 Hardware write protection activated I1650 Hardware write protection deactivated I1651 Custody transfer parameter changed...
  • Page 168 Diagnostics and troubleshooting Proline Cubemass C 300 Manufacturer →  168 Order code →  168 →  168 Extended order code 1 Extended order code 2 →  168 Extended order code 3 →  169 ENP version →  169 →...
  • Page 169 Proline Cubemass C 300 Diagnostics and troubleshooting Parameter Description User interface Factory setting Extended order code 3 Shows the 3rd part of the extended order Character string – code.  The extended order code can also be found on the nameplate of the sensor and transmitter in the "Ext.
  • Page 170 "Manufacturer' s information" document. The manufacturer' s information is available: • In the Download Area of the Endress+Hauser web site: www.endress.com → Downloads • Specify the following details: •...
  • Page 171 →  194. 13.2 Measuring and test equipment Endress+Hauser offers a wide variety of measuring and test equipment, such as W@M or device tests. Your Endress+Hauser Sales Center can provide detailed information on the services. List of some of the measuring and testing equipment: →  174→  176 13.3...
  • Page 172 • The measuring devices have a modular design. • Spare parts are grouped into logical kits with the associated Installation Instructions. • Repairs are carried out by Endress+Hauser Service or by appropriately trained customers. • Certified devices can only be converted to other certified devices by Endress+Hauser Service or at the factory.
  • Page 173 Proline Cubemass C 300 Repair 14.5 Disposal If required by the Directive 2012/19/EU on waste electrical and electronic equipment (WEEE), the product is marked with the depicted symbol in order to minimize the disposal of WEEE as unsorted municipal waste. Do not dispose of products bearing this marking as unsorted municipal waste.
  • Page 174 Various accessories, which can be ordered with the device or subsequently from Endress +Hauser, are available for the device. Detailed information on the order code in question is available from your local Endress+Hauser sales center or on the product page of the Endress+Hauser website: www.endress.com.
  • Page 175 Proline Cubemass C 300 Accessories 15.2 Communication-specific accessories Accessories Description Commubox FXA195 For intrinsically safe HART communication with FieldCare via the USB interface. HART  Technical Information TI00404F HART Loop Converter Is used to evaluate and convert dynamic HART process variables to analog current HMX50 signals or limit values.
  • Page 176 Accessory Description Applicator Software for selecting and sizing Endress+Hauser measuring devices: • Choice of measuring devices with industrial requirements • Calculation of all the necessary data for identifying the optimum flowmeter: e.g. nominal diameter, pressure loss, flow velocity and accuracy.
  • Page 177 Proline Cubemass C 300 Technical data Technical data 16.1 Application The measuring device is only suitable for the flow measurement of liquids and gases. Depending on the version ordered, the measuring device can also measure potentially explosive, flammable, poisonous and oxidizing media.
  • Page 178 Technical data Proline Cubemass C 300 16.3 Input Measured variable Direct measured variables • Mass flow • Density • Temperature Calculated measured variables • Volume flow • Corrected volume flow • Reference density Measuring range Measuring range for liquids Measuring range full scale values ...
  • Page 179 • Operating pressure to increase accuracy (Endress+Hauser recommends the use of a pressure measuring device for absolute pressure, e.g. Cerabar M or Cerabar S) •...
  • Page 180 Technical data Proline Cubemass C 300 Status input Maximum input values • DC –3 to 30 V • If status input is active (ON): R >3 kΩ Response time Configurable: 5 to 200 ms Input signal level • Low signal: DC –3 to +5 V •...
  • Page 181 Proline Cubemass C 300 Technical data 16.4 Output Output signal Current output 4 to 20 mA HART Order code "Output; input 1" (20): Option BA: current output 4 to 20 mA HART Signal mode Can be set to: • Active •...
  • Page 182 Technical data Proline Cubemass C 300 Damping Configurable: 0 to 999.9 s Assignable measured • Mass flow variables • Volume flow • Corrected volume flow • Density • Reference density • Temperature • Electronics temperature • Oscillation frequency 0 • Oscillation damping 0 •...
  • Page 183 Proline Cubemass C 300 Technical data Maximum output values 22.5 mA Maximum input voltage DC 30 V Load 0 to 700 Ω Resolution 0.38 µA Damping Configurable: 0 to 999 s Assignable measured • Mass flow variables • Volume flow •...
  • Page 184 Technical data Proline Cubemass C 300 Assignable measured • Mass flow variables • Volume flow • Corrected volume flow • Density • Reference density • Temperature • Electronics temperature • Oscillation frequency 0 • Oscillation damping 0 • Signal asymmetry •...
  • Page 185 Proline Cubemass C 300 Technical data Pulse/pause ratio Assignable measured • Mass flow variables • Volume flow • Corrected volume flow • Density • Reference density • Temperature  The range of options increases if the measuring device has one or more application packages.
  • Page 186 Technical data Proline Cubemass C 300 Current output 0/4 to 20 mA 4 to 20 mA Failure mode Choose from: • 4 to 20 mA in accordance with NAMUR recommendation NE 43 • 4 to 20 mA in accordance with US •...
  • Page 187 Proline Cubemass C 300 Technical data Interface/protocol • Via digital communication: HART protocol • Via service interface • CDI-RJ45 service interface • WLAN interface Plain text display With information on cause and remedial measures Web browser Plain text display With information on cause and remedial measures...
  • Page 188 Technical data Proline Cubemass C 300 Order code for Terminal voltage Frequency range "Power supply" DC 24 V ±20% – Option I AC 100 to 240 V –15 to +10% 50/60 Hz Power consumption Transmitter Max. 10 W (active power) switch-on current Max.
  • Page 189 Proline Cubemass C 300 Technical data 16.6 Performance characteristics Reference operating • Error limits based on ISO 11631 conditions • Water with +15 to +45 °C (+59 to +113 °F) at 2 to 6 bar (29 to 87 psi) • Specifications as per calibration protocol •...
  • Page 190 Technical data Proline Cubemass C 300 SI units 1:10 1:20 1:50 1:100 1:500 [mm] [kg/h] [kg/h] [kg/h] [kg/h] [kg/h] [kg/h] 0.04 22.5 1 000 US units 1:10 1:20 1:50 1:100 1:500 [inch] [lb/min] [lb/min] [lb/min] [lb/min] [lb/min] [lb/min] ¹⁄₂₄ 0.735 0.074...
  • Page 191 Proline Cubemass C 300 Technical data Influence of ambient Current output temperature Temperature coefficient Max. 1 μA/°C Pulse/frequency output Temperature coefficient No additional effect. Included in accuracy. Influence of medium Mass flow and volume flow temperature o.f.s. = of full scale value If there is a difference between the temperature during zero adjustment and the process temperature, the additional measured error of the sensors is typically ±0.0002 %o.f.s./°C...
  • Page 192 Technical data Proline Cubemass C 300 [% o.r./bar] [% o.r./psi] [mm] [in] ¹⁄₂₄ –0.001 –0.00007 ¹⁄₁₂ ¹⁄₈ –0.005 –0.0004 ¹⁄₄ –0.003 –0.0002 Design fundamentals o.r. = of reading, o.f.s. = of full scale value BaseAccu = base accuracy in % o.r., BaseRepeat = base repeatability in % o.r.
  • Page 193 4 to 95%. Operating height According to EN 61010-1 • ≤ 2 000 m (6 562 ft) • > 2 000 m (6 562 ft) with additional overvoltage protection (e.g. Endress+Hauser HAW Series) Degree of protection Transmitter • IP66/67, Type 4X enclosure, suitable for pollution degree 4 •...
  • Page 194 Technical data Proline Cubemass C 300 Options Oil- and grease-free version for wetted parts, without declaration Order code for "Service", option HA Mechanical load Transmitter housing: • Protect against mechanical effects, such as shock or impact • Do not use as a ladder or climbing aid...
  • Page 195 Proline Cubemass C 300 Technical data Seals For mounting sets with screwed-on connections: • Viton: –15 to +200 °C (–5 to +392 °F) • EPDM: –40 to +160 °C (–40 to +320 °F) • Silicone: –60 to +200 °C (–76 to +392 °F) •...
  • Page 196 Technical data Proline Cubemass C 300 Different values due to different transmitter versions: Transmitter version for the hazardous area (Order code for "Housing", option A "Aluminum, coated"; Ex d): +2 kg (+4.4 lbs) Weight in SI units Weight [kg] [mm]...
  • Page 197 Proline Cubemass C 300 Technical data Sensor housing • Acid and alkali-resistant outer surface • Stainless steel 1.4301 (304) Measuring tubes Stainless steel, 1.4539 (904L) Process connections VCO connection: VCO connection: stainless steel, 1.4539 (904L) Adapter for DN 15 flange according to EN 1092-1 (DIN2501) / according to ASME B 16.5 / as per JIS B2220:...
  • Page 198 Technical data Proline Cubemass C 300 Surface roughness All data refer to parts in contact with the medium. The following surface roughness categories can be ordered. Not polished 16.11 Operability Languages Can be operated in the following languages: • Via local operation...
  • Page 199 Proline Cubemass C 300 Technical data Via remote display and operating module DKX001 The remote display and operating module DKX001 is available as an optional extra →  174. • The measuring device is always supplied with a dummy cover when the remote display and operating module DKX001 is ordered directly with the measuring device.
  • Page 200 Technical data Proline Cubemass C 300 Supported operating tools Different operating tools can be used for local or remote access to the measuring device. Depending on the operating tool used, access is possible with different operating units and via a variety of interfaces.
  • Page 201 Proline Cubemass C 300 Technical data Supported functions Data exchange between the operating unit (such as a notebook for example) and the measuring device: • Upload the configuration from the measuring device (XML format, configuration backup) • Save the configuration to the measuring device (XML format, restore configuration) •...
  • Page 202 The device meets the legal requirements of the applicable EU Directives. These are listed in the corresponding EU Declaration of Conformity along with the standards applied. Endress+Hauser confirms successful testing of the device by affixing to it the CE mark. UKCA marking The device meets the legal requirements of the applicable UK regulations (Statutory Instruments).
  • Page 203 Proline Cubemass C 300 Technical data Contact address Endress+Hauser UK: Endress+Hauser Ltd. Floats Road Manchester M23 9NF United Kingdom www.uk.endress.com RCM mark The measuring system meets the EMC requirements of the "Australian Communications and Media Authority (ACMA)". Ex approval The devices are certified for use in hazardous areas and the relevant safety instructions are provided in the separate "Safety Instructions"...
  • Page 204 The application packages can be ordered with the device or subsequently from Endress+Hauser. Detailed information on the order code in question is available from your local Endress+Hauser sales center or on the product page of the Endress+Hauser website: www.endress.com.
  • Page 205 Proline Cubemass C 300 Technical data Data logging (line recorder): • Memory capacity for up to 1000 measured values is activated. • 250 measured values can be output via each of the 4 memory channels. The recording interval can be defined and configured by the user.
  • Page 206 For an overview of the scope of the associated Technical Documentation, refer to the following: • Device Viewer (www.endress.com/deviceviewer): Enter serial number from nameplate. • Endress+Hauser Operations app: Enter serial number from nameplate or scan matrix code on nameplate. Standard documentation Brief Operating Instructions...
  • Page 207 Proline Cubemass C 300 Technical data Supplementary device- Safety instructions dependent documentation Safety instructions for electrical equipment for hazardous areas. Contents Documentation code ATEX/IECEx Ex d/Ex de XA01480D ATEX/IECEx Ex ec XA01481D cCSAus XP XA01482D cCSAus Ex d/ Ex de...
  • Page 208 Technical data Proline Cubemass C 300 Installation Instructions Contents Comment Installation instructions for spare part sets and • Access the overview of all the available spare part sets via Device Viewer →  172 accessories • Accessories available for order with Installation Instructions →  174...
  • Page 209 Proline Cubemass C 300 Index Index Connection see Electrical connection Access authorization to parameters Connection preparations ..... . 33 Read access ......55 Connection tools .
  • Page 210 Index Proline Cubemass C 300 Display and operating module DKX001 ..199 Field Communicator 475 ..... . 69...
  • Page 211 Proline Cubemass C 300 Index Installation conditions Mounting requirements Dimensions ......22 Down pipe .
  • Page 212 Index Proline Cubemass C 300 Current output ......87 Read access ....... . 55 Current output (Wizard) .
  • Page 213 Proline Cubemass C 300 Index Totalizer reset ......140 Web server ......62 WLAN .
  • Page 214 Index Proline Cubemass C 300 Use of the measuring device Borderline cases ......9 Incorrect use .
  • Page 216 *71574975* 71574975 www.addresses.endress.com...