GE MEDICAL SYSTEMS
REV 21
SECTION (E) Preset all the parameters reset by the initialization.
11. Preset all the parameters reset by the initialization.
11–1.Hospital Name setting.
i. Press the <Control>, H and then the <Enter> key.
ii Type "Hospital Name" and press the <Enter> key. Confirm that the Hospital Name is displayed in
upper–left corner on the monitor.
iii Press the <Control> key.
11–2.Record Output (Gamma Curve Selection) setting.
i. Preset the Record Output (Gamma Curve Selection) recorded in step 2–2.
Refer to the Operator Manual to preset the Record Output (Gamma Curve Selection). If user was using
polaroid , preset the Film Development Time.
Film Development Time setting.
i. Preset the Film Development Time recorded in step 2–2.
Refer to the Operator Manual to preset the Film Development Time.
11–3.Inverse setting.
i. Preset the Inverse setting recorded in step 2–3
Refer to the Operator Manual to preset the Inverse.
11–4.Map Package setting.
Press the <Control> key
.
11–5.Dotted Line setting.
i. Preset the Dotted Line setting recorded in step 2–5.
Refer to the Operator Manual to preset the Dotted Line.
11–6.Frame Averaging setting.
i. Preset the Frame Averaging setting recorded in step 2–6.
Refer to the Operator Manual to preset the Frame Averaging.
11–7.OB Table Data settings.
i. Preset the OB Table Data recorded in step 2–7.
Refer to the Operator Manual to preset OB Table Data.
11–8.Body Pattern Package settings.
i. Preset the Body Pattern Package recorded in step 2–8.
Refer to the Operator Manual to preset the Body Pattern Package.
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUAL
Yes
Press the <Control> key
3–51
Press N key
No
Is the display the
same as step 2–4?
Press the <Enter> key
INSTALLATION AND REPLACEMENT
P9030GX