3–7–3 Chassis Leakage Current Test (Continued)

Data Sheet for Chassis Source Leakage Current

Follow the foregoing test procedure. The test passes when all readings measure less than the value shown in the
table below (TABLE 3–3).

MAXIMUM ALLOWANCE LIMIT FOR CHASSIS SOURCE LEAKAGE CURRENT

COUNTRY
USA
OTHERS
Electric Shock Hazard. When the meter's ground switch is OPEN, don't touch the unit!
Unit
Tester
Power
POLARITY
Switch
Enter name of tested peripheral here:
ON
NORM
ON
NORM
ON
REV
ON
REV
OFF
NORM
OFF
NORM
OFF
REV
OFF
REV
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUAL
TABLE 3–3
NORMAL
OPEN
CONDITION
GROUND
N/A
0.1mA
0.1mA
0.5mA
CAUTION
TABLE 3–4
DATA SHEET FOR CHASSIS SOURCE LEAKAGE CURRENT
Tester
Test1
NEUTRAL
Probe
Switch
connector
OPEN
CLOSED
OPEN
CLOSED
OPEN
CLOSED
OPEN
CLOSED
3–12
REVERSE
OPEN
POLARITY
NEUTRAL
0.1mA
N/A
0.5mA
0.5mA
Test2
Test3
Test4
wheel
CRT
INSTALLATION AND REPLACEMENT