GE MEDICAL SYSTEMS
REV 23
1–2–1 Warnings (Continued)
The following table describes the purpose and location of safety labels and other important information provided on the
equipment.
Label/Symbol
Identification and Rating
Plate
Type/Class Label
Device Listing/
Certification Labels
"DANGER – Risk of explosion
used in..."
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUAL
Purpose/Meaning
Manufacturer's name and address
Date of manufacture
Model and serial numbers
Electrical ratings
Used to indicate the degree of safety
or protection.
Equipment Type BF (man in the box
symbol) IEC 878–02–03 indicates B
Type equipment having a floating ap-
plied part.
Laboratory logo or labels denoting
conformance with industry safety
standards such as UL or IEC.
The system is not designed for use
with flammable anesthetic gases.
"ATTENTION – Consult accompa-
nying documents " is intended to alert
the user to refer to the operator
manual or other instructions when
complete information cannot be pro-
vided on the label.
"Mains OFF" Indicates the power
off position of the mains power
switch.
"Mains ON" Indicates the power
on position of the mains power
switch.
"Protective Earth" Indicates the
protective earth (grounding) terminal.
1–8
Location
Rear of console near power
inlet
Probe connectors
and PCG connector
Rear of console
Rear of console
Various
Adjacent to On–Off Switch
Adjacent to On–Off Switch
Various
INTRODUCTION
P9030GX