Sulfate - Hanna Instruments HI83325 Instruction Manual

Multiparameter photometer
Hide thumbs Also See for HI83325:
Table of Contents

Advertisement

11.10. SULFATE

SPECIFICATIONS
Range
Resolution
Accuracy
Light Source
Method
REQUIRED REAGENTS
Code
HI93751-0
REAGENT SETS
HI93751-01
HI93751-03
For other accessories see ACCESSORIES section.
MEASUREMENT PROCEDURE
• Select the
• Fill a cuvette with 10 mL of unreacted sample (up to the mark).
Replace the plastic stopper and the cap.
• Insert the cuvette into the holder and close the lid.
• Press Zero. The display will show "-0.0-" when the meter is zeroed and ready for measurement.
66
66
66
0 to 150 mg/L (as SO
1 mg/L
±5 mg/L ±3% of reading at 25 °C
LED with narrow band interference filter @ 466 nm
Sulfate is precipitated with barium chloride crystals
Description
Sulfate Reagent
Reagents for 100 tests
Reagents for 300 tests
Sulfate
method using the procedure described in the METHOD SELECTION section.
2-
)
4
Quantity
1 packet
10 mL