11.6. MAGNESIUM

SPECIFICATIONS
Range
0 to 150 mg/L (as Mg
Resolution
1 mg/L
Accuracy
±5 mg/L ±3% of reading at 25 °C
Light Source
LED with narrow band interference filter @ 466 nm
Method
Adaptation of the Calmagite Method
REQUIRED REAGENTS
Code
Description
HI93752A-Mg
Magnesium Reagent A
HI93752B-Mg
Magnesium Reagent B
REAGENT SETS
HI937520-01
Reagents for 50 tests
HI937520-03
Reagents for 150 tests
For other accessories see ACCESSORIES section.
MEASUREMENT PROCEDURE
• Select the
Magnesium
method using the procedure described in the METHOD SELECTION section.
• Add 1 mL of
HI93752A-Mg
Reagent A to the cuvette using a 1 mL
syringe and use the pipette to fill the
cuvette up to the 10 mL mark with the
HI93752B-Mg
Magnesium Reagent B.
• Replace the plastic stopper and the cap.
Invert several times to mix.
• Insert the cuvette into the holder and close
the lid.
2+
)
Magnesium
Quantity
1 mL
9 mL
1 mL
10 mL