Hanna Instruments HI83325 Instruction Manual page 52

Multiparameter photometer
Hide thumbs Also See for HI83325:
Table of Contents

Advertisement

MEASUREMENT PROCEDURE
• Select the
• Fill the cuvette with 10 mL of unreacted sample (up to the mark).
Replace the plastic stopper and the cap.
• Insert the cuvette into the holder and close the lid.
• Press Zero. The display will show "-0.0-"; the meter is zeroed and ready for measurement.
• Remove the cuvette and add the content of one packet of
HI96777A-0
stopper and the cap. Shake gently for 30 seconds.
• Insert the cuvette into the holder and close the lid.
52
52
52
Iron(II)/(III)
method using the procedure described in the METHOD SELECTION section.
Iron(II)/(III) Reagent A. Replace the plastic
10 mL