4. Configuration
OK, High, and Low can be set to NO (normally open, Type A) or NC (normally closed, Type B).
The default value is NO (normally open, Type A)
 OK, High, and Low are NO (normally open, Type A)  default
 OK, High, and Low are NC (normally close, Type B)
Make J3, J4, and J5 open circuit by cutting the circuit
Then make J6, J7, and J8 short circuit with solder.
Restore factory settings when RJ45 option card is installed
If scale does not work properly after modifying setting through RJ45 card, please follow the
steps below to restore to factory setting.
In the weighing mode, hold the
second). Then release key, and "RELOAD" appears. When reloading end, please restart the
scale to restore to the factory settings.
02036XXX
CN6
key while press
23
key at the same time (within 1
ZSME300000422