Purpose; Tools And Expendables Required; Parts Required; Documents Required - Siemens MAMMOMAT 3000 Modular Modification Instructions

Installation of the option w/rh anode/filter combination
Table of Contents

Advertisement

Prerequisites

Purpose

The purpose of this modification is to upgrade a MAMMOMAT 3000 Modular with Mo
X-ray tube to Mo/W X-ray tube.

Tools and expendables required

Standard assembly tools
Circlip pliers
Denatured ethanol, 99.5 % pure (may only be denatured with toluene, which does not
leave any residues when evaporated) for cleaning the filters or tube window.
Lint-free cloth or Q-tips (cotton tops) for cleaning
Lint-free cotton gloves
Service PC with connecting cable (99 00 440 RE999)

Parts required

Filter-assembly kit, part no. 64 82 967 X041E, comprising the following parts:
Qty
1
1
5
4
1

Documents required

Installation and Start-Up Instructions for MAMMOMAT 3000 Modular.

Time required

Approximately four hours for one person.

Important notes

The Allen screws can be very hard to loosen. When unscrewing, be sure of enough grip
with the Allen screwdriver, so that the screws do not get damaged.
The filters in the filter assembly must not be contaminated in any way. Therefore, always
use lint-free cotton gloves during the handling of the filter assembly. Do not remove the
filter assembly from its packing until installation.
If the filters have been contaminated, clean by using the specified alcohol. Other types
of solvents may leave residues.
Siemens-Elema AB
Solna, Sweden
Part
Filter assembly
Retaining ring
Allen screw M5 (countersunk)
Cable tie
Adapter cable
Register 5
SPB7-230.092.11
Rev. 01
04.98
1
Page 1 of 2
MAMMOMAT 3000 Modular
SPS-UD
1 - 1
1
1
1
1
1
1
Service

Advertisement

Table of Contents
loading

Table of Contents